CytoTherm CT-DR Скачать руководство пользователя страница 9

 

SOURCES for SUPPLIES 

BAGS overwrap 15 cm. wide by 30 cm. (6” x 12”) long are available from US Plastics 1390 Neubrecht 
Rd. Lima OH 45801 Tel. 800 537-9724. Part #47352. 

 

TECHNICIAN FUNCTIONS 

 
FULL CALIBRATION 
 
Adjust  the  10-turn  potentiometer  so  that  the  voltage  between  TP1  and  Ground  (Earth)  is  .395  V  (see  wiring 
diagram  for  locations).  Actual  temperature  calibration  is  then  set  in  Programming  Mode  (see 
PROGRAMMING MODE). 
The trimmer near the BODY connector adjusts the display contrast. 
 

SAFETY FEATURES 

 
WATER LEVEL SENSOR 
The unit will display LOW WATER and beep if there is not enough water to cover the level sensor inside the 
bath, a screw located 3/4" above the temperature sensor. The heat will also turn off.  The water must be 
conductive in order for the level sensor to work. Use tap water or add a pinch (0.01 gr.) of salt if you are using 
de-ionized or distilled water. 
 
OVER TEMPERATURE 
The unit will display TOO HOT and emit double-beeps if the temperature reaches 1.0

o

C above the thawing 

temperature. The controls will turn the heaters off.  In the unlikely situation that there is a component failure 
and the unit continues heating, an  independent safety thermostat will turn the heating off at 45 

o

C.  

 
SAFETY THERMOSTAT 
An independent safety thermostat will turn heaters off if the temperature goes above 45 

o

C. It has to be cooled 

to 20 C to reset. 
 
FUSING 
The main fuses (Fuse-1 & Fuse-2) are located on the Printed Circuit Board and are 8Amp for 120Vac and 
5Amp for 240Vac. Correct the cause of the blown fuse. 
 
The transformer has a ¾ Amp slow blow fuse (Fuse-3) on the wires of the 12 Vac winding used to generate the 
5 Vdc supply. 
 
ISOLATION 
The circuitry is isolated from the power lines by a transformer and from the heaters by an opto-coupled solid 
state relay. 

 

 
 
 
 
 
 

Содержание CT-DR

Страница 1: ...well Ave Trenton NJ USA Tel 609 396 1456 Fax 609 396 9395 Do not forget to mail in your warranty card Please record the following Serial No ___________ Date of Purchase _________ TABLE OF CONTENTS Dra...

Страница 2: ...ION The notched side of the ARMS face up Place the BATH in front of the CONTROL SECTION so that the chassis on the BATH is next to the indented portion of the CONTROL SECTION Connect the umbilical cor...

Страница 3: ...hawed on the bottom part of the BLADDER Cover the plasma bag s with the top layer of the BLADDER Tuck the bladder in Close the LID by lifting straight up until it stops and then rotating it down forwa...

Страница 4: ...leach solution Remount a new BLADDER You are ready to start thawing again Rinse off the bleach with water 3 Plasma bag and BLADDER both leaked It is a very unlikely case After you determine that the p...

Страница 5: ...is 1 0C above the target temperature 5 Select the signal when thawing is done continuous or pulsed Enter PROGRAM MODE by holding ENTER while turning the unit ON Follow the displayed instructions to pr...

Страница 6: ...Was the unit filled with water over 40C If yes then cool the water to 20C to reset the safety thermostat Check the Safety thermostat C TH and the heaters 14 77 for continuity Bag leaked but does not...

Страница 7: ...nd displayed alarm that indicates that a plasma bag or BLADDER leaked MAINTENANCE Daily Wipe off the unit Every 3 Months Clean the BATH Unplug the unit Remove and discard the BLADDERS by first lifting...

Страница 8: ...IBRATION Overtemperature Alarm Test Hold MINUS as you turn the unit ON DIAGNOSTICS will be displayed Push ENTER Push and release PLUS until ALARM TEST is displayed Push ENTER the heater will heat cont...

Страница 9: ...water or add a pinch 0 01 gr of salt if you are using de ionized or distilled water OVER TEMPERATURE The unit will display TOO HOT and emit double beeps if the temperature reaches 1 0o C above the th...

Страница 10: ...Power Cord Orn Blu Orn Blu Grn Yel Yel Mag Sensor Sensor Connector Level Sensor in Rack Level Sensor in Rack Start Stop Minus Plus Enter Reed Sensor Rock Position Thermistor Blu Grn Vio 2Grn 2Orn 2Br...

Отзывы: