RINSING the BATH. Rinse off the bleach with water. Close the DRAIN by removing the Drain Tubing. Twist 
counterclockwise lightly and pull out. Fill the bath with water to 2.5 Cm. (1 Inch) above the pivot screws. If you 
use deionized or distilled water add a little pinch ( about 0.01 gr.) of table salt to make the water conductive. 
 
Drain the BATH a second time. Refill with water and connect the pump tubing to the RACK. Add 5.5 ml. of 
Algaecide Part #ALG240 (1.0 ml. per 4 liters of water) . Available USA only) to the water.  Replace the old 
bladders with new bladders.   
 
Replace the air filter.  Unscrew the filter cover, discard the old filter and mount the new filter. Screw on the 
filter cover. 
 
Leak Test  Turn the unit ON. Take a slightly moist piece of a paper towel and touch it to both of the stainless 
steel  studs  near  the  center  of  each  RACK  section  under  the  BLADDER..  The  alarm  should  sound  and  the 
display should say LEAK. If it does not, contact CytoTherm.   
 
Temperature  Calibration.  Run  temperature  calibration  at  least  every  3  months  or  more  often  if  your 
institution procedures or regulations require. See PROGRAM MODE > TEMPERATURE CALIBRATION. 
 
Overtemperature Alarm Test.  Hold MINUS as you turn the unit ON.  DIAGNOSTICS will be displayed. Push 
ENTER. Push and release PLUS until  ALARM TEST is displayed.  Push ENTER, the heater will heat 
continuously and the pumps will turn on. When the temperature reaches 1.0C above the target temperature 
the alarm will sound. Turn Off the unit.  If alarm does not sound, contact CytoTherm. Alarm test should be 
performed every 3 months as part of unit maintenance. 
 
 
BLADDER Maintenance. The bladder does not require maintenance. In case of a plasma bag leak, remove the 
BLADDER,  rinse  the  BLADDER  under  running  water  and  wipe  with  bleach  towelettes  or  a  10%  bleach 
solution.    
 

PARTS LIST CT-DR

 

 

CDR-MP 

Pump, Magnetic 

DP-SW   

Switch, 2-pole    

14-77 

Heater (2 needed) 

C-TH 

Safety Thermostat 

CT4-PC 

Printed Circuit Board 

C4-DISP 

Display 

CDR-BTH 

Bath Section 

C4-UMB 

Wiring Harness 

C-RM 

Rocking Motor 

CDR-PM 

Program Module 

C4-TS 

Temperature Sensor 

CDR-SCN 

Sensor Connector 

CDR-RK 

Rack for Dry Thaw 

CDR-LID 

Lid for Rack 

CDR-BL 

Bladder (3 needed) 

C4-TR 

Transformer 120 Volt 

C4-TRJ 

Dual Transformer. Japan only 

C4-TR2 

Transformer 240 Volt 

ALG240 

Algicide 

C4-FILT 

Air Filters,   per pack of 4  

C-VLVF 

Drain Valve Female  

Mounted on Bath 

C-VLVM  

Drain Valve Male 

With tubing 

14-77J 

Heater 100V (use 2) Japan 

 

 

Always specify model and serial number when ordering. 
 
 
 

Содержание CT-DR

Страница 1: ...well Ave Trenton NJ USA Tel 609 396 1456 Fax 609 396 9395 Do not forget to mail in your warranty card Please record the following Serial No ___________ Date of Purchase _________ TABLE OF CONTENTS Dra...

Страница 2: ...ION The notched side of the ARMS face up Place the BATH in front of the CONTROL SECTION so that the chassis on the BATH is next to the indented portion of the CONTROL SECTION Connect the umbilical cor...

Страница 3: ...hawed on the bottom part of the BLADDER Cover the plasma bag s with the top layer of the BLADDER Tuck the bladder in Close the LID by lifting straight up until it stops and then rotating it down forwa...

Страница 4: ...leach solution Remount a new BLADDER You are ready to start thawing again Rinse off the bleach with water 3 Plasma bag and BLADDER both leaked It is a very unlikely case After you determine that the p...

Страница 5: ...is 1 0C above the target temperature 5 Select the signal when thawing is done continuous or pulsed Enter PROGRAM MODE by holding ENTER while turning the unit ON Follow the displayed instructions to pr...

Страница 6: ...Was the unit filled with water over 40C If yes then cool the water to 20C to reset the safety thermostat Check the Safety thermostat C TH and the heaters 14 77 for continuity Bag leaked but does not...

Страница 7: ...nd displayed alarm that indicates that a plasma bag or BLADDER leaked MAINTENANCE Daily Wipe off the unit Every 3 Months Clean the BATH Unplug the unit Remove and discard the BLADDERS by first lifting...

Страница 8: ...IBRATION Overtemperature Alarm Test Hold MINUS as you turn the unit ON DIAGNOSTICS will be displayed Push ENTER Push and release PLUS until ALARM TEST is displayed Push ENTER the heater will heat cont...

Страница 9: ...water or add a pinch 0 01 gr of salt if you are using de ionized or distilled water OVER TEMPERATURE The unit will display TOO HOT and emit double beeps if the temperature reaches 1 0o C above the th...

Страница 10: ...Power Cord Orn Blu Orn Blu Grn Yel Yel Mag Sensor Sensor Connector Level Sensor in Rack Level Sensor in Rack Start Stop Minus Plus Enter Reed Sensor Rock Position Thermistor Blu Grn Vio 2Grn 2Orn 2Br...

Отзывы: