MODES 

 

STANDBY MODE.  Just turn the unit ON. The unit heats up and maintains 37.0

O

C in the BATH.   

 
THAWING MODE. The unit recirculates tempered water through the bladders, rocks the BATH and RACK, times the 
thawing and monitors for leaks. Pushing START initiates thawing. Pushing STOP stops thawing. 
 
DIAGNOSTIC MODE. Hold MINUS as you turn the unit ON. DIAGNOSTICS is displayed.  

1.

 

To restore the unit to initial factory settings, while still holding MINUS press ENTER. The unit displays “UNIT 
RESET”. Turn the unit off. You have to re-calibrate the temperature after resetting the unit. 

2.

 

To proceed to diagnostics without resetting the unit, release MINUS and press ENTER. Push and release PLUS to 
advance to the different tests. 

3.

 

“PUMP TEST” is a diagnostic tool to check if  the pump and rocker motor turn on. The display shows “PUMP 
TEST”. MINUS turns on the Pump, ENTER turns the rocking motor. Push PLUS to go to next function. 

4.

 

Overtemperature Alarm Test.  Hold MINUS as you turn the unit ON.  DIAGNOSTICS will be displayed. 
Push ENTER. Push and release PLUS until  ALARM TEST is displayed.  Push ENTER, the heater will 
heat continuously and the pumps will turn on. When the temperature reaches 1.0C above the target 
temperature the alarm will sound. Turn Off the unit.  If alarm does not sound, contact CytoTherm. 
Alarm test should be performed every 3 months as part of unit maintenance.

 

 
PROGRAM MODE.  Allows you to: 
  1.     Program the thawing time. 
  2.     Select whether to continue thawing when timed out.   

3.

 

Calibrate the temperature. 

4.

 

Set the target thawing temperature. The Alarm temperature is 1.0C above the target temperature. 

5.

 

Select the signal when thawing is done (continuous or pulsed). 

 
Enter PROGRAM MODE by holding ENTER while turning the unit ON. Follow the displayed instructions to program. 
Pushing PLUS advances the menu to the next parameter.  
 
Programming the Thawing Time. When the thawing time is displayed, press MINUS because you want to change the time. 
Use PLUS or MINUS to set the desired new thawing time, press ENTER to program in the new time. Turn the unit off.  
 
Calibrating Temperature. Turn the unit on and let it heat up to 37.0

o

C.  Check the displayed temperature with your 

accurate thermometer. There is a covered THERMOMETER WELL on the right side of the RACK. It is covered with a 
cover bar. Rotate the cover bar to expose the thermometer well. Place your thermometer into the THERMOMETER 
WELL. The thermometer stem should be at least 13 cm. (5”) long. If the displayed temperature needs to be changed, turn 
the unit off.  Enter PROGRAMMING MODE by holding ENTER as you turn the unit on. When the thawing time is 
displayed push PLUS  until the display shows “CALIBRATE TEMPERATURE”. Push MINUS because you want to 
change the displayed temperature. Use PLUS or MINUS to make the displayed temperature agree with your 
thermometer. Push ENTER to record the calibration. Turn the unit off.  If the unit can not be calibrated, or if  the 
calibrated temperature is not stable,  have a technician check the unit. See TECHNICIAN FUNCTIONS.  
 

SWITCH  FUNCTIONS   

There are 6 switches on the front panel:

 

START.  Starts function. 
STOP.  Stops function. 
ENTER.  Records parameter changes. 
PLUS.  Steps through parameters and increases value of parameter. 
MINUS.  Allows change in parameter and decreases value of parameter.  
Sketch of FIN logo.  Not currently used.  Future expansion

 

Содержание CT-DR

Страница 1: ...well Ave Trenton NJ USA Tel 609 396 1456 Fax 609 396 9395 Do not forget to mail in your warranty card Please record the following Serial No ___________ Date of Purchase _________ TABLE OF CONTENTS Dra...

Страница 2: ...ION The notched side of the ARMS face up Place the BATH in front of the CONTROL SECTION so that the chassis on the BATH is next to the indented portion of the CONTROL SECTION Connect the umbilical cor...

Страница 3: ...hawed on the bottom part of the BLADDER Cover the plasma bag s with the top layer of the BLADDER Tuck the bladder in Close the LID by lifting straight up until it stops and then rotating it down forwa...

Страница 4: ...leach solution Remount a new BLADDER You are ready to start thawing again Rinse off the bleach with water 3 Plasma bag and BLADDER both leaked It is a very unlikely case After you determine that the p...

Страница 5: ...is 1 0C above the target temperature 5 Select the signal when thawing is done continuous or pulsed Enter PROGRAM MODE by holding ENTER while turning the unit ON Follow the displayed instructions to pr...

Страница 6: ...Was the unit filled with water over 40C If yes then cool the water to 20C to reset the safety thermostat Check the Safety thermostat C TH and the heaters 14 77 for continuity Bag leaked but does not...

Страница 7: ...nd displayed alarm that indicates that a plasma bag or BLADDER leaked MAINTENANCE Daily Wipe off the unit Every 3 Months Clean the BATH Unplug the unit Remove and discard the BLADDERS by first lifting...

Страница 8: ...IBRATION Overtemperature Alarm Test Hold MINUS as you turn the unit ON DIAGNOSTICS will be displayed Push ENTER Push and release PLUS until ALARM TEST is displayed Push ENTER the heater will heat cont...

Страница 9: ...water or add a pinch 0 01 gr of salt if you are using de ionized or distilled water OVER TEMPERATURE The unit will display TOO HOT and emit double beeps if the temperature reaches 1 0o C above the th...

Страница 10: ...Power Cord Orn Blu Orn Blu Grn Yel Yel Mag Sensor Sensor Connector Level Sensor in Rack Level Sensor in Rack Start Stop Minus Plus Enter Reed Sensor Rock Position Thermistor Blu Grn Vio 2Grn 2Orn 2Br...