Connect tubing
Follow the instructions in the table below to connect the tubing from ÄKTA avant and
ÄKTA pure to Fraction collector F9-R, or to both Fraction collector F9-R and Fraction
collector F9-R, 2nd.
Step
Action
1
Lift out the Tubing holder (4) from the Delivery arm (1).
3
4
5
6
Part
Function
1
Delivery arm
2
Tube adjustment cavity
3
Sensor control
4
Tubing holder
5
Tubing holder nut
6
Exposed tubing end
2
Loosen the nut of the Tubing holder. Do not remove the Tubing holder nut (5)
from the Tubing holder.
3
Insert the tubing through the Tubing holder.
ÄKTA avant and ÄKTA pure Instructions 29021462 AD