Cygnus DIVE Operating Manual
M3-DIVE-M-ENG_Iss9.docx
Page 41 of 117
nearby surfaces, causing the echo bars to increase - this is
normal.
Stability Indication in Single Echo & Echo-Echo
Modes
To help indicate when a Single Echo or
Echo-Echo measurement is stable
–
and
thus probably reliable
–
the gauge changes
the colour of the thickness measurement
number from Red to Yellow when the
measurement has been stable for 2
consecutive seconds.
When measuring using SE or EE mode
once you have a measurement keep the
probe still and wait for the gauge to signal
a “stable reading”.
If the ultrasound signal is poor or erratic
then the thickness value may remain Red
–
thus indicating the measurement may
not be reliable.
When using the Data Logging feature
–
measurement will only be logged when the
gauge has detected a stable reading.
Not Stable
–
RED
Stable - Yellow