M5-CYG4-M-01_Iss6.doc
Cygnus 4 Operating Manual
29
2.
Place the Probe on the sample
and verify the gauge can get a
thickness value.
3.
Access the Menu and scroll down
to the ‘Calibration’ group. Then
scroll right to the ‘1 Point Cal’
item.
Press the Ok/Select key to open
the ‘1 Point Cal’ function
4.
While holding the probe firmly
on the sample, and while a
steady thickness measurement
is displayed..
5.
...use the Up and Down keys to
adjust the thickness to the
required value.
6.
When done press the OK/Select
key to save the calibration. A
Calibration Saved message will
be shown.
Or press the X/Cancel key to exit
without saving the calibration.
When a Calibration has been made a ‘1PC’ indictor is displayed at
the top of the measurement screen in the status area (1PC =
Single Point Calibration).