6.1
Chapter 6
Specifications
System Requirements
To run the WaferSense ATS software and link, your computer must have:
•
Windows 2000, Windows XP, or Windows Vista operating system
•
One free high-power USB 1.1 or USB 2.0 port
ATS Hardware
Environmental
Operating pressure range:
≤
10
-6
Torr to atmospheric.
Storage temperature range: -20 °C to 70 °C.
Charging temperature range: 0 °C to 45 °C
Operating temperature range: Optimum 20 °C to 50 °C. Can withstand exposure to 120 °C
(5 minutes or less) if not in direct contact with heating element.
Power
Battery charger requires 100 - 240 VAC at 47 - 63 Hz input.
Battery usage on a full charge: approximately 2 hours. Battery performance degrades at
temperatures outside the optimum operating temperature range.
Typical battery recharge cycles: approximately 500.
Range and Accuracy
X and Y accuracy: ±0.1 mm (±0.004")
Z range: 6.4 mm to 45 mm (0.25" to 1.8")
Z accuracy: ±0.5 mm (±0.02")