41
EXCEL HR OPERATOR MANUAL
D1796, REV. D, 12/16
3.
Clean the proximal surface of the cooling window, as follows:
Prior to treatment, clean the proximal surface of the cooling window with a germicidal
disposable wipe.
During treatment, keep the proximal surface of the cooling window clear of gel by gently
wiping periodically with 4x4 gauze. Excessive top side gel may affect laser beam output,
possibly leading to adverse events such as burns.
The proximal surface of the cooling window has a delicate optical
coating; take care to not scratch this surface. Use a new germicidal
disposable wipe or 4x4 gauze, as applicable, each time the proximal
surface is cleaned.
Proximal surface
Patient contact
surface
4.
Inspect the cooling window. If necessary, clean the window again.
The cooling window must be clean before treating a patient. Chips,
pits, cracks or burned spots on the window are not acceptable, and
a handpiece with these types of defects must be replaced.
NOTE
CAUTION