5
Page
CALIBRA
TION CHECK PROCEDURE
QUICK START GUIDE
STANDARD
MICROMETER
•
Remove reservoir plug and replace with Oil Fill and Vent
Cap. The Oil Fill and Vent cap and the work light bulb are found in
the accessories box.
•
Place the crimper on a sturdy bench in a well lit area, and
plug the crimper directly into a 115 V 15 Amp wall outlet.
Do not
run the crimper on an extension cord as low voltage can damage
the motor.
•
The oil level in the pump should be approximately
1-1/2 inches below the
fi
ll plug. If necessary,
replenish with ISO Viscosity Grade 46 hydraulic
oil.
Note: The crimper is calibrated prior to shipment, but
a calibration check is recommended prior to using
the crimper for the
fi
rst time.
•
Lubricate all contact surfaces and place
the Cone Insert, any Small Series die set and
the Standard Pusher in the crimper in the order
shown. Note that a hose and
fi
tting are not required for
an initial calibration check.
Note: failure to lubricate all
contact surfaces can cause the Cone Insert and/or the dies
to seize in the cone base.
•
Slide the Pusher onto the stud on the hydraulic ram.
•
Set the Micro-Crimp adjuster at “95”
•
Press and hold the start switch bulb.
•
If the ram extends and the crimper shuts off approximately
1 second after the dies are completely closed and the pump starts
to build pressure, the crimper is correctly calibrated.
•
If the time to shut off is not approximately 1 second, the
crimper must be recalibrated. Recalibrate per instructions in this
manual
FOLLOW THESE STEPS BEFORE USING THE CRIMPER FOR THE FIRST TIME