Ergometer ec3000
Interfaces
11-3
11.3. Remote interface
The device can be connected to an external computer system with a RS 232-C serial
interface. The following functions can be carried out:
Remote
Execution of pc-supported ergometry programs
Acknowledgement of ergometry data
The external computer system can be connected as required to a DIN-round connector or
a Sub-D-connector. The interface is recognized automatically.
Safety instruction:
Only one of the two interfaces must be connected.
The following table describes the major characteristics of the system:
Function
Specification
Notes
Baud rate
2400 Bd ( Custo Control )
4800 Bd ( P10 )
Data format
8 Bit / no parity /
1 Stopbit
Protocol
Custo Control / P10
cf. interface specification
Isolation
4 kV
cf. DIN EN 60601-1-1