7.
Adjust focus manually or using
AutoFocus
. Autofocus works depending on the AF set range so perform this
job after adjusting AF with Jog
(page 17).
8. On the
Imaging control panel
, adjust
LED intensity, brightness and contrast
by using scroll bar.
✓
Lamp power
:
Adjustable from 0 to 255
✓
Brightness
: Adjustable from -2048 to 2047
✓
Contrast
: Adjustable from 0 to 160
Note
: The recommended values are as follows.
Lamp power
: 135,
Brightness
: 256,
Contrast
: 40. The higher the
difference between cell and background, the better calculation of confluency
(page 33).
9. Select
Image Format
and
Frame Resolution
in
Imaging
.
Frame Resolution
can also be selected on the left side of the
display
.
10. To save the image, right-click on the display and click Save As. The saved images can be seen in ScreenCaptured file.
Two image files, one with raw image and the other one with feature are generated.
<Raw image>
<
Image with feature
>