14
2. Cell monitoring
[
Positioning (position selection) and light intensity control]
1.
Put the device in the incubator.
2.
When the stage comes out by pressing , install the vessel. (In case of dishes or flasks, install the vessel
adapter first, then the vessel. Refer to Appx B for how to insert a vessel or vessel adapter
(page 32).
Upon installing
well plate, make sure that A1 is on the left front.
3.
Select the vessel type in
Setup
mode. Well plate, dish, flask can be selected. Upon selecting vessel type, the vessel
type is displayed on the plate panel at the bottom right.
Note
: A vessel adapter suitable for vessel type must be used. (use vessel adapter for vessels except for well plate)
4.
Click and move desired vessel location on the plate panel with the mouse. (Upon clicking vessel or well in Plate
panel, it moves to center of vessel or well).
5.
Upon clicking
Start Streaming
, a real-time image is shown on the display. [If the screen is black, light intensity
should be adjusted first (
step 8
)].
Plate panel
A1