5.2.4
CLEANING THE APPLIED PARTS
CAUTION
Danger of material damages. Only use detergent and disinfecting
products specifically approved for medical devices or medical-
surgical devices.
Applied parts that come into direct contact with the patient during
the examination shall be thoroughly cleaned after each use with a
disinfectant approved for the purpose.
1
Turn off the device and unplug it from the electrical outlet.
2
Clean the applied parts using products suitable for surface
disinfection (they may contain aldehyde).
Alternatively, use a non-abrasive cloth soaked in a solution of
water, ethyl alcohol (70% maximum) or isopropyl alcohol.
For more information about suitable cleansing products, read the
“Recommended products for cleaning and disinfection” paragraph
on page 49.
5.2.5
CLEANING THE OPTICAL COMPONENTS
CAUTION
Danger of material damages. The device is equipped with optical
components. The optical components of the device are precision
and pressure sensitive parts. Clean using a non-abrasive cloth to
avoid damaging the surface.
Clean the optical components carefully using a dry, non-abrasive, lint-
free cloth.
5.3
DEVICE CALIBRATION
Perform device calibration periodically, in order to ensure accurate
measurements. Follow the calibration instructions given in paragraph
“How to perform device calibration” on page