INSTRUCTIONS FOR USE
| IFU327EN00 - 01/2021
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
30
/56
The accessories must be compliant to the norm IEC 60950-1
Information technology equipment - Safety - Part 1: General
requirements.
If the accessories are installed in the patient area it is necessary to
install an insulating transformer compliant with the directive IEC
60601-1:2005 + A1:2102 - "Medical electrical equipment - Part 1:
General requirements for basic safety and essential performance".
3.2
TECHNICAL DATA
Technical data
Value
Data transfer
USB 3.0
Mains power
External power supply unit 24 VCC. In:
100-240Vac - 50/60Hz - 0.9-05A Out:
24Vdc - 40W
Net cable
with C14 socket
Dimensions
(Height x Length x Depth)
515 x 315 x 255 mm
Weight
7 kg
Chin rest stroke
70 mm ±1
Minimum height of the chin cup from
the working plan
24 cm
Base movement (x, y, z)
105 x 110 x 30 mm
Working distance
74 mm
Light sources
Technical data
Value
Placido
LED @400 - 700 nm
Scheimpflug
LED @475 nm UV-free
Pupillography
LED @940 nm
Lighting for Fluorescein exam
LED @470 nm
Auxiliary lighting
LED @400 - 700 nm