PC: 4 GB RAM - Video Card 1 GB RAM (not shared) resolution 1024
x 768 pixels
-
Operating system: Windows XP, Windows 7 and Windows 10
(32/64 bit).
Read the instructions for use of the application software.
The PC must be compliant to the norm IEC 60950-1 Information
technology equipment - Safety - Part 1: General requirements.
If the PC is installed in the patient area it is necessary to install an
isolation electrical supply compliant with the directive IEC 60601-
1:2005 + A1:2012- "Medical electrical equipment - Part 1: General
requirements for basic safety and essential performance".
It is possible to connect other accessories to the PC (printer, modem,
scanner, etc) through the analogical or digital interfaces.
The accessories (printer, modem, scanner, etc) must be installed
outside the patient area.
The accessories must be compliant to the norm IEC 60950-1
Information technology equipment - Safety - Part 1: General
requirements.
If the accessories are installed in the patient area it is necessary to
install an isolation electrical supply compliant with the directive IEC
60601-1:2005 + A1:2102 - "Medical electrical equipment - Part 1:
General requirements for basic safety and essential performance".