Controller Messages
STATUS
MESSAGE
ACTION
MESSAGE
MEANING WHEN DISPLAYED
USER ACTION
PUMP OFF
START PUMP
Pump is off with no notifications
Start pump
PUMP ON
SET SPEED
Pump is on, no speed selected
Select speed
LOW SPEED
NONE
Low speed selected
N/A
MEDIUM SPEED
NONE
Medium speed selected
N/A
HIGH SPEED
NONE
High speed selected
N/A
SPEED OUT OF
RANGE
CALL SERVICE
Speed is detected, when stop has been
requested
Return controller for
service
PUMP FAILURE
CALL SERVICE
DEFECTIVE PUMP
Return controller for
service
PRESSURE OUT
OF RANGE
CALL SERVICE
Pressure is detected, when stop has been
requested
Return controller for
service
FLUID SUPPLY
EMPTY
FLUID SUPPLY
EMPTY
Controller has calculated that the saline
bag is empty
Change saline bag and
reset
MAX PRESSURE
EXCEEDED
MAX PRESSURE
EXCEEDED
The controller has detected that output
pressure exceeded maximum for > 10
seconds
Reselect lower speed or
exchange device
MAX SPEED
EXCEEDED
MAX SPEED
EXCEEDED
The controller has detected that output
speed exceeded maximum for > 10
seconds
Reselect lower speed or
exchange device
NO SPEED
DETECTED
CHECK AND
RETRY
No Tachometer reading detected
1. Check fiber optic
connections
2. Replace fiber optic
SYSTEM FAILURE
CALL SERVICE
Controller Failure
Return controller for
service
PUMP STOPPED
CLOSE DOOR
Saline pump is open
Close pump
NO SPEED
DETECTED
CHECK AND
RETRY
Speed test failed
1. Check device
connections
2. Replace device
SALINE LOW
CHECK BAG
Controller has calculated a low saline
supply
Check saline level and
change bag. Reset level
when needed.
EVENT TIME
EXCEEDED
NONE
Event Time has exceeded 90 seconds
Follow event times
listed in OAD IFU
APPROACHING
MAX EVENT TIME
NONE
Event Time is approaching 90 seconds
Follow event times
listed in OAD IFU
FLOW OUT OF
RANGE
NONE
Controller has detected improper saline
flow rate condition.
Call for Assistance