8
©
2009 Cardiovascular Systems, Inc.
90074-00.F
Procedure
Refer to OAD Instructions for Use for procedural information.
Saline Bag Change
When the Saline Low notification appears it is time to change the saline bag.
1. Stop the pump by touching the pump On/Off button.
2. Touch the Change IV Bag Button
3. Remove the bag spike from the bag and hold it for reuse.
4. Remove the empty bag from the pole.
5. Place a new bag on the pole and insert the bag spike.
6. Touch the OK button in the operation window to reset the saline volume to full.
a. If saline bag has not been changed touch Cancel in the operation window
7. Restart the pump by touching the Turn Pump On button in the operation window.
Device Change
To change out a device:
1. Stop the pump by touching the pump On/Off button.
2. Disconnect the tubing from the device handle in use.
3. Attach the tubing to the new device.
4. Touch the Turn Pump On button in the operation window to start the pump to prime saline through the
device. Stepping on the foot pedal will increase the flow rate to high for faster priming
a. Note: Controller will default to Test Mode (60kRPM) once pump has been turned on.
5. Refer to OAD Instructions for Use for procedural information.
Turning Off the Controller
1. Stop the pump by touching the pump On/Off button.
2. Touch the Turn System Off button in the operation window
3. Touch OK to shut down system
4. When the CSI splash screen is seen turn off the main power to the controller by pressing the On/Off button on
the back of the controller.
Cleaning the Controller and Foot Pedal
The Controller System should be cleaned regularly with a soft cloth dampened in water and mild detergent. Isopropyl
alcohol/ water mix (70/30) or Cydex should be used to clean and disinfect the Controller System from blood or other
contaminants. Do not immerse the Controller, Monitor or Foot Pedal in fluid. Avoid use of solvents or abrasive
cleaners, which may damage the Controller System components.
Controller Maintenance and Calibration
The Controller does not require routine or periodic maintenance or calibration. Contact CSI if questions arise related
to Controller function or performance.
Controller Repair and Shipping
Contact CSI if Controller repairs are required or for shipping instructions.
Classifications, per IEC/UL/CSA/EN 60601-1
•
Class I Equipment
•
Type CF Applied Part
•
Degree of protection against the ingress of water: IPX1 (Controller), IPX0 (Monitor), IPX8 (Footswitch)
•
Equipment not suitable for use in the presence of flammable anesthetics
•
Operation Mode: Continuous
Storage,Transport, and Operating Conditions
Storage and Transport: -50
°
C to +60
°
C (-58
°
F to 140
°
F), 10% to 100% Relative Humidity, 5000 to 1060 hPA
Operating: 10 to 40
°
C (50 to 104
°
F)
Electromagnetic Compliance
Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to
be installed and put into service according to the EMC information provided.
a. Portable and mobile radio frequency (RF) communications equipment can affect medical electrical
equipment. This Product is intended for use in the electromagnetic environments specified.
b. The end user of this product should assure it is used in such an environment.
i. Portable and mobile RF Communications equipment should not be used at close distances.
ii. Power frequency magnetic fields should be at levels characteristic of a typical location in a typical
commercial or hospital environment.