Appendix E.
Tera Term
QuickStart
E-9
13.
After the reprogramming is successful, disconnect power from the sensor
and disconnect the serial cable from the DB-9 connector.
14.
Move the jumper from
PROG
back to
RUN
.
15.
Reassemble the sensor.
a.
Line up the notch on the disk/circuit assembly with the V in the main
metal housing. Do not kink the wires going from the disk/circuit
assembly to the main metal assembly. Seat the disk/circuit assembly.
Put the two Phillips screws back in the countersunk holes. Do not over
tighten the screws. They should be snug.
b.
Older transducer heads use a removable O-ring. Carefully seat the
O-ring back into place and make sure you won’t cut it when
assembling the transducer assembly.
c.
Line up the connector on the transducer housing with the mating
connector on the disk/circuit assembly and slide it into place until the
tab clicks into place.
d.
Line up the notch on the transducer housing with the V on the main
metal housing. Seat the housing.
e.
Put the six screws back into the bottom of the transducer and tighten
them down.
16.
Verify operation of the sensor by obtaining a measurement.