Appendix D. RS-232 and RS-485 Operation
D-8
The frequency that the SR50A performs the measurement is set by adjusting
the
Measurement Interval Units
and the
Measurement Interval Value
parameters.
•
The main advantage of this operating mode is that the receiving device
will only have to wait 100 ms for the data instead of 1 second.
•
This configuration is also more conducive to a multidrop RS-485 system
where individual sensors do not transmit until they are addressed.
D.4.4 Distance to Target or Depth
The SR50A can output either distance to target values or calculate snow depth
values. To obtain a valid snow depth value the parameter distance to ground
must be entered in.
The SR50AT will compensate the readings for temperature.
Do not use this option on the SR50A sensor unless the SR50A is sent valid
temperature reading via the
Temperature Input
command.
D.4.5 Distance to Ground
A valid distance to ground must be entered when the SR50A is configured to
output snow depth values. The value must be in meters regardless of the output
units that are selected.
If the exact value cannot be obtained, it is better to slightly overestimate the
value rather than underestimating it. If a
Distance to Ground
value is too
small, the SR50A will output an error value as the snow surface should not be
below the ground surface.
D.4.6 Measurement Interval Units
This setting is only applicable if either the
Auto Measure Polled Output
or the
Auto Measure Auto Output
modes are used. The options for the
Measurement
Interval Units
are:
•
Seconds
•
Minutes
•
Hours
Once a unit type is selected, the number of units for the interval is set by
changing the
Measurement Interval Value
parameter. A 60 s interval can be
set by setting the units to seconds and the
Measurement Interval Value
to 60.
Alternately, the
Measurement Interval Unit
could be set to minutes and the
Value
could be set to 1. The
Value
setting can only range from 1 to 255.
D.4.7 Measurement Interval Value
This setting is only applicable if either the
Auto Measure Polled Output
or the
Auto Measure Auto Output
modes are used. The
Measurement Interval Value
can range from 1 to 255. The units used for the value is set by the
Measurement Interval Units
.
Содержание SR50A
Страница 2: ......
Страница 6: ......
Страница 30: ...SR50A Series Sonic Ranging Sensors 20...
Страница 32: ......
Страница 53: ...Appendix B Example Programs B 21 CallTable Daily CallTable Hour NextScan EndProg...
Страница 54: ...Appendix B Example Programs B 22...
Страница 56: ......
Страница 78: ...Appendix E Tera Term QuickStart E 10...
Страница 83: ......