C-1
Appendix C. Jumper Settings
shows the jumper locations of the SR50A.
FIGURE C-1. Jumper Settings
The SR50A can be configured with either SDI-12, RS-232 or RS-485
communications. Shunt jumpers can be located on any of the three sets of
communication selection headers. Only place the three jumpers on one group at
a time. Never install more than three jumpers and never mix the jumpers
among the SDI-12, RS-232 or RS-485 locations.
The other jumper located on the SR50A places the sensor in either the normal
operation mode or in the program update mode. The program mode is only
used for updating the internal firmware of the sensor. During operation, have
the jumper in the
RUN
position.
Refer to Appendix E.2,
Sending New Firmware to a SR50A/T Sensor
for details
about SR50A
firmware updates
.