FIGURE 9-3. Proper location of the gas analyzer top wick (left) and
bottom wick (right)
9.3 Gas-Analyzer Window Cleaning
The windows of the analyzer should be cleaned if the signal strength for CO
2
or H
2
O drops below 0.7 or 70% of the original value. These values may be
monitored in the output data, or they can be viewed with
ECMon
. To clean the
windows, use isopropyl alcohol (2-propanol / isopropanol) and a cotton swab
or a non-scratching tissue or cloth. Verify that following cleaning the signal
strengths are restored to values close to 1.0.
9.4 Zero and Span
As is the case with all optical instruments, the IRGASON’s gas analyzer
measurements may drift slightly with exposure to natural elements. Therefore,
routine maintenance requires a periodic zero-and-span procedure. The
frequency with which this should be done is heavily dependent on the
environmental conditions in which the instrument is making measurements.
Campbell Scientific recommends that a new instrument, or an instrument that
has been moved to a new location, should have a monthly zero and span
procedure performed. Additionally, it is recommended to always zero and span
the analyzer after cleaning the windows (see Section
(p. 28)
). After the first several zero-and-span procedures, the
rate of drift in gain and offset (explained later in this section) will be better
understood for the existing environmental conditions. A regimented protocol
for zero and span of the instrument initially, will allow the user to assess the
appropriate time interval between zero and span procedures. However, if the
IRGASON is being used at a site with large seasonal changes in ambient
conditions, the zero-and-span procedure should be done at least seasonally.
In situations in which spanning the instrument is extremely difficult (for
example, in below freezing temperature conditions or in very remote sites),
performing a monthly zero will help mitigate the drift in the analyzer’s
measurements.
Содержание IRGASON
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 12: ...Table of Contents iv ...
Страница 60: ...Appendix A Filter Bandwidth and Time Delay A 4 ...
Страница 64: ...Appendix C Material Safety Data Sheets MSDS C 2 ...
Страница 65: ...Appendix C Material Safety Data Sheets MSDS C 3 ...
Страница 66: ...Appendix C Material Safety Data Sheets MSDS C 4 ...
Страница 67: ...Appendix C Material Safety Data Sheets MSDS C 5 ...
Страница 68: ...Appendix C Material Safety Data Sheets MSDS C 6 ...
Страница 69: ...Appendix C Material Safety Data Sheets MSDS C 7 ...
Страница 70: ...Appendix C Material Safety Data Sheets MSDS C 8 C 2 Decarbite MSDS ...
Страница 71: ...Appendix C Material Safety Data Sheets MSDS C 9 ...
Страница 72: ...Appendix C Material Safety Data Sheets MSDS C 10 ...
Страница 73: ......