Sensor-
Settings / Type „Modbus“
PI 500
Seite
53
von
93
Example:
Holding Register - UI1(8b) - Value: 18
Holding Register
– UI4(32) - Value: 29235175522
à
AE41 5652
Main menu
Settings
Sensor settings
C1
Unit- description field
Selection Register Type
Holding Register
,
Data Type
U1(8b
) und Byte Order
A / B
HByte LByte
18 => 00 12
Data Order 1. Byte 2. Byte
A
00
12
B
12
00
Selection Register Type
Holding Register
,
Data Type
U1(32b
) und Byte Order
A-B-C-D
HWord
LWord
HByte LByte HByte LByte
29235175522 => AE 41 56 52
Data Order 1.Byte 2.Byte 3.byte 4.Byte
A-B-C-D AE 41 56 52
D-C-B-A 52 56 41 AE
B-A-D-C 41 AE 52 56
C-D-A-B 56 52 AE 41
By pressing the description field
Unit
the list with the available units appear
Please select the unit by pressing the
respective button e.g.
m³/h.
For validation of the unit, please push the
button
OK
To move through the list please press the
button
Page.
In case the unit is
not
available it is possible
to create a user defined unit.
Therefore, please select one of the
User_X
buttons.