background image

CS Instruments 

- 4 - 

Air leaks produce wide-band ultrasounds in the range of 20 ... 80 kHz. The 
higher the frequency the more energy it contains. But higher frequencies 
can't be transported in air that far. That's why LD 300 operates at a centre 
frequency of 40 kHz which compromises an optimum between energy and 
distance. Frequencies below and above are cut in order to minimize the 
noise level. 

 

Scope of delivery 

 

Description 

Order No. 

LD 300 Leak Detector set consisting of: 

0601 0103 

 

LD 300 Leak Detector 

0560 0102 

 

Sensor unit 

0605 0001 

 

Noise isolated head set 

 

0554 0102 

 

Focus tube and focus tip 

0530 0101 

 

Cable to detach sound probe from instrument 

0553 0101 

 

Battery charger 

0554 0001 

 

Transport case 

0554 0101 

Accessories not includes in set: 

 

Ultrasonic Tone Generator 

0554 0103 

Telescope extension with cable 

0530 0102 

Содержание LD 300

Страница 1: ...INSTRUCTION MANUAL V2 01 2007 LD 300 Leak Detector For leak detection in pneumatic systems...

Страница 2: ...on compliance with this manual Should the device be tampered with in any manner other than a procedure which is described and specified in the manual the warranty is cancelled and the manufacturer is...

Страница 3: ...l to confirm a diagnosis on the spot by being able to clearly discriminate among various equipment sounds This is accomplished by an electronic process called heterodyning that accurately converts the...

Страница 4: ...istance Frequencies below and above are cut in order to minimize the noise level Scope of delivery Description Order No LD 300 Leak Detector set consisting of 0601 0103 LD 300 Leak Detector 0560 0102...

Страница 5: ...on about 4 hours with LASER pointer on Charging time around 1 5 hour Charging operation temp 10 C to 45 C recommended Operating temperature 0 C to 40 C Storage temperature 10 C to 50 C Weight 2 5 kg...

Страница 6: ...sec all LCD segments are on LD 300 is switched off LD 300 is switched on LCD all segments on for 2 sec Laser off Press Press Press after 30 minutes the appliance switches off The user can adjust the s...

Страница 7: ...With all operating equipment most leakage and electrical problems produce a broad range of sound The high frequency ultrasonic components of these sounds are in nature extremely short wave A short wa...

Страница 8: ...s at insulations Insulation test on doors and windows gas leaks in general Leak detection in compressed air systems Compressed air is one of the most expensive energy forms Only in Germany 60 000 pneu...

Страница 9: ...ose up detection the sensor unit can be removed from the main unit In the scope of delivery there is an extension cable which can be connected between sensor and main unit Detection of leaks in unpres...

Страница 10: ...is is also the case if the serial number has been changed damaged or removed The warranty time for the LD 300 is 12 months for the instrument 6 months for accessory parts Warranty services do not exte...

Страница 11: ...ssories CS Instruments GmbH as the manufacturer herewith declares that the above instruments and accessories belongs to the category 9 WEEE 2002 96 EC Therefore the above instruments do not fall upwar...

Страница 12: ...uth Zindelsteiner Str 15 D 78052 VS Tannheim Tel 49 0 7705 97 89 9 0 Fax 49 0 7705 97 89 9 20 info cs instruments com www cs instruments com Sales office North Am Oxer 28c D 24955 Harrislee Tel 49 0 4...

Отзывы: