INSTALLATION

Requirements for installation in hazardous area

Zone 1

Humidity / Dew point transmitter FA 300-2 Ex is only allowed to be connected to an intrinsic safe
circuit, in which an electrically isolated supply and signal circuit in the power supply is preferred.
Equipment without electrical isolation according to EN 50020 must not be used, if a potential
equalization occurs along the wire. The related requirements according to EN 60079-14 have to
be met.

Uo = 30 V max.
Io = 100 mA max.
Po = 1  W  max.

The maximum effective inner capacity is 17 nF, the maximum effective inner inductance is neglec-
table.

4 to 20 mA analogue output

Order No. 0700.3021

FA 300-2 Ex

Connection of transmitter to ATEX certified supply

Explosive area

non explosive area

Barrier

Power supply

PLC

Содержание FA 300-2 Ex

Страница 1: ...tationary humidity measuring instrument for measuring pressure dew point and atmospheric dewpoint for explosive areas FA 300 2 Ex INSTRUCTION MANUAL Process safety for compressed air systems V1_04_200...

Страница 2: ...r compressed air 6 Installation 7 8 Calibrating Adjusting Connection 9 Ordering data 10 Warranty 11 Contact 12 Cost performance ratio Reliable quality at a fair price We have the ideal solutions for y...

Страница 3: ...Lager und Transporttemperatur sowie die zul ssige Betriebstemperatur beachten z B Messger t vor direkter Sonneneinstrahlung sch tzen Bei ffnen des Ger ts unsachgem er Behandlung oder Gewaltanwendung...

Страница 4: ...4 Dimensions in mm DIAGRAM OF INSTRUMENT FA 300 2 Ex Cable max 8...

Страница 5: ...Operating temperature 20 to 70 C Storage temperature 40 to 80 C Load for analogue output 500 Ohm Screw in thread G 1 2 stainless steel Material of housing polycarbonate Sensor protection sinter filter...

Страница 6: ...may result in a reduction of the pressure dew point value Pressure dew point C Gramme of water vapour per m 3 of moisture saturated compresse air g m 3 Air pressure bar Air pressure bar with referenc...

Страница 7: ...not use ordinary plastic tubes The dew point temperature in Ctd does not change when cooled as it is an absolute humidity value which like other measured variables e g g m3 is independent of temperat...

Страница 8: ...t electrical isolation according to EN 50020 must not be used if a potential equalization occurs along the wire The related requirements according to EN 60079 14 have to be met Uo 30 V max Io 100 mA m...

Страница 9: ...scheme We recommend a yearly cycle If requested we can carry out calibration on your premises Connection Basic instruments FA 300 2 Ex 4 to 20 mA analogue output in 2 wire technology Opening the plug...

Страница 10: ...re dewpoint measuring instrument 80 to 20 Ctd 0699 3070 Special scaling analog output in g kg rF mg m3 0699 4004 Precision calibration at 40 Ctd incl certificate 0699 3396 Intrinsically safe power sup...

Страница 11: ...WARRANTY If in addition to the warranty service necessary repairs adjustments or similar are carried out the warranty services are free of charge but there is a charge for other services such as tran...

Страница 12: ...8052 VS Tannheim Tel 49 0 7705 97899 0 Fax 49 0 7705 97899 20 info cs instruments com www cs instruments com Am Oxer 28c D 24955 Harrislee Tel 49 0 461 7002025 Fax 49 0 461 7002026 info cs instruments...