INSTALLATION

In general humidity can be measured in all non-corrosive

active gases. For corrosive gases please query with

CS INSTRUMENTS GmbH.

To enable accurate measurements in the low dew point range (–30 to –80 °C

td

), the measuring

temperature of the gas should, if possible, be that of room temperature (20 to 35 °C). With resin
driers, for example, or other applications, the temperature of the measuring gas is often higher,
e.g. 70 ° to 120 °C.
In this case we recommend installing a "cooling tunnel" of impermeable material in front of the
screw-on measuring chamber.
A Teflon pipe or a copper pipe would be ideally suitable for this purpose, as the hot gas is 
cooled to ambient temperature over the length of the pipe, approx. 2 to 5 m.

Please do not use ordinary plastic tubes!

The dew point temperature in °C

td

does not change when cooled as it is an absolute humidity

value, which, like other measured variables e.g. g/m

3

, is independent of temperature.

Druckluftleitung

FA 300

I N S T A L L A T I O N

FA 300-2 Ex

Compressed air pipe

Directly in the compressed air system

Screw in probe with G ½ " thread pressure-tight in the cen-
tre or in the compressed air pipe where the measurement is
to take place.Ensure that the measurement is carried out
close to the compressed air flow. U-bend pipes or non-flow-
ing compressed air result in very slow reaction times for the
moisture reading. Installation is recommended following dry-
ing of the compressed air and all bypass pipes or for critical
compressed air users.

I N S T A L L A T I O N

Measurable gases:

I N S T A L L A T I O N

Consideration: 

When installing the instrument in areas

of gas group IIC ensure, that there are no sparks even
in uncommon hazardous incidents.

Содержание FA 300-2 Ex

Страница 1: ...tationary humidity measuring instrument for measuring pressure dew point and atmospheric dewpoint for explosive areas FA 300 2 Ex INSTRUCTION MANUAL Process safety for compressed air systems V1_04_200...

Страница 2: ...r compressed air 6 Installation 7 8 Calibrating Adjusting Connection 9 Ordering data 10 Warranty 11 Contact 12 Cost performance ratio Reliable quality at a fair price We have the ideal solutions for y...

Страница 3: ...Lager und Transporttemperatur sowie die zul ssige Betriebstemperatur beachten z B Messger t vor direkter Sonneneinstrahlung sch tzen Bei ffnen des Ger ts unsachgem er Behandlung oder Gewaltanwendung...

Страница 4: ...4 Dimensions in mm DIAGRAM OF INSTRUMENT FA 300 2 Ex Cable max 8...

Страница 5: ...Operating temperature 20 to 70 C Storage temperature 40 to 80 C Load for analogue output 500 Ohm Screw in thread G 1 2 stainless steel Material of housing polycarbonate Sensor protection sinter filter...

Страница 6: ...may result in a reduction of the pressure dew point value Pressure dew point C Gramme of water vapour per m 3 of moisture saturated compresse air g m 3 Air pressure bar Air pressure bar with referenc...

Страница 7: ...not use ordinary plastic tubes The dew point temperature in Ctd does not change when cooled as it is an absolute humidity value which like other measured variables e g g m3 is independent of temperat...

Страница 8: ...t electrical isolation according to EN 50020 must not be used if a potential equalization occurs along the wire The related requirements according to EN 60079 14 have to be met Uo 30 V max Io 100 mA m...

Страница 9: ...scheme We recommend a yearly cycle If requested we can carry out calibration on your premises Connection Basic instruments FA 300 2 Ex 4 to 20 mA analogue output in 2 wire technology Opening the plug...

Страница 10: ...re dewpoint measuring instrument 80 to 20 Ctd 0699 3070 Special scaling analog output in g kg rF mg m3 0699 4004 Precision calibration at 40 Ctd incl certificate 0699 3396 Intrinsically safe power sup...

Страница 11: ...WARRANTY If in addition to the warranty service necessary repairs adjustments or similar are carried out the warranty services are free of charge but there is a charge for other services such as tran...

Страница 12: ...8052 VS Tannheim Tel 49 0 7705 97899 0 Fax 49 0 7705 97899 20 info cs instruments com www cs instruments com Am Oxer 28c D 24955 Harrislee Tel 49 0 461 7002025 Fax 49 0 461 7002026 info cs instruments...

Отзывы: