CoCo-80 Quick Reference
71
Issue
Possible Cause
Solution
detected by
EDM.
installed.
CD under
User Guide and Tutorial>
Technical Support Documents
, and
finally open the
How To Setup the
CoCo USB Driver
tutorial.
Antivirus
software is
preventing
data transfer.
•
Check the settings of the
antivirus software and give
permission for CoCo
connections.
•
Check with local IT
department to allow the
connection.
Physical
connection
error.
•
Check the connection status
on the CoCo. In the top
ribbon of the CoCo display is
a computer icon with either
a red X or a green check. A
red X indicates that a
physical connection does not
exist.
•
If using an Ethernet
connection verify that both
the host PC and the CoCo are
both set to either Dynamic IP
settings or both are set to
similar Static IP settings.
Sensor
input is
lower than
expected.
Check that the
sensor channel
has been set
correctly.
•
Verify
Measurement
Quantity
has been set
correctly.
•
Verify
Sensor Sensitivity
is
correct.