2.3 Bump test
If the Gas-Pro is configured for Bump/Pump functionality, then by placing the Gas-Pro in a Q-Test
module, the Bump/Pump screen will be displayed (see
i
If the Gas-Pro is a non pumped unit and not configured for bump test the ‘sensors
blocked by flow plate’ warning icon will be shown when inserted into the Q-Test. The Gas-
Pro will need to be configured for bump to operate correctly in the Q-Test.
i
If the Gas-Pro is a pumped unit and not configured for bump test the pump will
immediately start. The Gas-Pro will need to be configured for bump to operate correctly in
the Q-Test.
Figure 8: Bump/Pump screen
Click the operator button to highlight
for Bump testing and then double click it to select the Bump test.
According to the Gas-Pro configuration, the user may be asked to choose which group of gases to
bump and either a Speedy bump or Smart bump test will be performed (see