Looking after Gas-Tec 

Gas-Tec

28

X Looking after Gas-Tec

10.1 Gas cylinder 

Start with a filled gas cylinder . The ‘gas pressure low’ prompt will show when the pressure is too low for 

the unit to function properly . 
A full 400 ml gas cylinder (150 bar, 2150 psi) should last for approximately 60 hours . The average hourly 

pressure drop is approximately 2 .5 bar (35 psi) .
Make sure the cylinder is turned off using the gas cylinder ON/OFF valve after each use .
See also Section XII Hydrogen Safety (p32)

10.2 Filters

The purpose of the filter assembly is to protect the flame chamber from contamination . If water or solid 

particles enter the chamber, it may cease to function properly . The accuracy of readings may be affected . 
The filter assembly consists of a filter backing disc, a 25 mm diameter hydrophobic filter element, an 'O' 

ring, and a 27 mm diameter paper pre-filter element . Assemble the filter using components in order as 

shown in the illustration below .

The  paper  pre-filter  should  be  inspected  regularly  and  replaced  when  necessary .  This  will  protect  the 

hydrophobic filter from coarse contamination and extend its life . When fitting this filter, the hydrophobic 

surface should be placed in contact the 'O' ring . The fibrous surface, marked with a red dot, should be 

placed face down against the backing disc .
If water enters the filter assembly, the hydrophobic surface will prevent moisture entering the flame cham-

ber, but before continuing to use the Gas-Tec, make sure the probe and pre-filters are clear .
Discard the pre-filter . 
Shake or blow lightly on the hydrophobic surface . 
Reassemble the unit with a new pre-filter .
Dry and clean the probe and tube assemblies thoroughly before reusing .
NOTE: Instruments should not be run without the filter assembly .

10.3 Probes and tubes 

Check probes and tubes for damage before use .
The water trap cavity at the bottom of the insulated probe handle 

should be cleaned out with a clean rag or tissue (do not use volatile 

solvents) . The joint should be opened periodically and any water 

shaken out . Avoid water passing into the instrument .

Содержание Gas-Tec

Страница 1: ...Gas Tec User manual M07979 Issue 1 October 2009...

Страница 2: ......

Страница 3: ...eading 18 5 2 Setting time and date 19 VI What the Gas Tec shows 20 6 1 Overview 20 6 2 Special events in the record 21 6 3 GPS 21 6 4 Acquiring and using data from the Gas Tec 23 VII Managing the Gas...

Страница 4: ...stru ment at 15 ml minute In extreme cases a build up could cause an explosion WARNING USE OF HYDROGEN As with any apparatus using hydrogen stored under high pressure certain precautions are prudent T...

Страница 5: ...handle Filter attachment Angled coat hanger probe Skids for coat hanger probe 2 Multiregional power supply 6 Spare filters 3 Manual 7 Strap 4 Quick Start Guide 8 Cylinder Cylinder not supplied in USA...

Страница 6: ...Overview Gas Tec 6 Diagram 1 1 1 ON OFF button 2 Battery indication Light GREEN Power OK Light RED Power LOW Flashing GREEN Charging 3 GPS antenna 4 Strap loops 4 1 8 Unit orientation 2 3...

Страница 7: ...Gas Tec Overview 7 10 11 9 12 5 Headphone jack 6 Power supply DC jack 7 USB port 8 Trigger jack 9 Probe connector 10 Gas cylinder on off valve 11 Igniter 12 Gas cylinder Diagram 2 7 5 6 8...

Страница 8: ...e battery is holding The LED to the left of the icon shows green for Power OK red for Power Low and flashes green while the unit is charging B The data icon shows the status of the storage medium C Th...

Страница 9: ...the concentration of hydrocarbons present in the sample All sam ples are destroyed as the measurements are generated This process takes place in a specially designed chamber 2 3 Hydrogen Safety Users...

Страница 10: ...hydrocarbon gases are present the Gas Tec reading will reflect their presence The Gas Tec s response to other gases differs from the reading shown for methane Correction factors for some common hydroc...

Страница 11: ...CD display these requirements can be checked once the machine has been switched on However there are a few steps to complete prior to this Attach the gas cylinder Set the hydrogen cylinder in place by...

Страница 12: ...tones 3 2 Start Up When the hydrogen cylinder and the probe are properly attached the Gas Tec can be switched on 1 Press the ON OFF button The LCD display will run a self test and then be ready for u...

Страница 13: ...l button under the Gas Tec Press it in until it clicks This should light the hydrogen and the beeping will stop If the beeping continues press the igniter again 9 When the gas is lit the flame icon wi...

Страница 14: ...the Gas Tec for invehicle surveys it is recommended that a lock is achieved outside of the vehicle before installation to the Trigger survey unit see Section VIII 2 takes place 3 The Gas Tec monitors...

Страница 15: ...tomatically the beep will stop and the message will clear If the Gas Tec falls sideways the flame may go out Check for this by inspecting the flame icon in the middle of the status bar at the top of t...

Страница 16: ...isable Pressure units psi bar Speed mph miles per hour km h kilometres per hour m s metres per second Kts knots Admin menu see Chapter VII IMPORTANT NOTE Accepting the option RESET DEFAULTS on this me...

Страница 17: ...nalogue Graph and analogue Display Backlight timer 30 seconds 1 minute 3 minute Disabled Time and date Set time Set date Set format Units Pressure units PSI Bar Speed MPH KM H M S KTS Restore Defaults...

Страница 18: ...bout how the data is displayed use the SetUp menu To do this switch the Gas Tec on and press the SETUP soft key 5 1 Choosing how to display the gas reading Press ENTER when the Format option is highli...

Страница 19: ...y timestamp is correct by using the Time and Date menu Choose to display these in a number of different ways Similarly there are options for the units to be used when the unit displays pressure and fo...

Страница 20: ...same time To upload the data plug the detector s USB connector into a PC the green led will light on the Gas Tec when connected Gas Tec will be recognised as a memory device and a folder will open dis...

Страница 21: ...GPS Read the details on the screen above right If GPS data has been recorded the data can be converted by web based applications and then displayed on a map Gas Tec thus provides a straightforward wa...

Страница 22: ...input form googleearth To convert the data saved from the Gas Tec onto your PC Follow the instructions in the screenshot below to process the data and create a file that Google Earth can open and disp...

Страница 23: ...The file name con tains the date of the survey Drop the files to a folder on the PC Alternatively cut and paste them Delete the copied files from the internal drive of the Gas Tec to free up space for...

Страница 24: ...tton down at the right hand side of the LCD display After entering the Admin menu the Gas Tec will not detect gas To use the Gas Tec turn it off using the on off button and then press again to turn it...

Страница 25: ...e first step Choose the last option to allow an Error Override this allows the Gas Tec to restart attempting to clear any known failure modes If the Gas Tec still reports an error after running this o...

Страница 26: ...heeled along over a smooth surface collecting samples at ground level Bell housing probe E for bore hole applications 2 Telescopic Probe with polytetrafluroethylene PTFE tubing for monitoring volume o...

Страница 27: ...e Lithium ion Users may not change the battery of this product it must be replaced at Crowcon or a service centre Battery life 22 hrs Recharge time 2 5 hrs Operating temperature range 5 C to 55 C 23 F...

Страница 28: ...lustration below The paper pre filter should be inspected regularly and replaced when necessary This will protect the hydrophobic filter from coarse contamination and extend its life When fitting this...

Страница 29: ...bar on the display will give five hours of operation The symbol will flash when the battery needs recharging Recharge the battery after each use of the Gas Tec so it will be ready for the next session...

Страница 30: ...193 Probe filter assembly kit FIT99034 Angled probe skids M01197 Probe converter kit converts old probes for new Gas Tec PRB64001 Straight probe attachment PRB99007 Probe cup attachment PRB99006 Bore...

Страница 31: ...brane and fixings S012986 Display PCB assembly M041041 Left and right blue sides includes fixings label E011055 Igniter assembly M04120 Tubing set comprises full kit for refit of 1 Gas Tec E011059 Bat...

Страница 32: ...as cylinders should never be used if pressure has not been reduced by a suitable regu lator at the cylinder or at the outlet of the header valve of a cylinder manifold Use only regulators intended to...

Страница 33: ...ressure and or replace the cylinder III x System Primed Prepared for firing Wait for new messages III x Auto Zero Place probe in clean Air Device prepared to set the zero PPM level Make sure that the...

Страница 34: ...Charger is required to continue Plug in the charger and power on the charger X 4 Lost GPS Signal No GPS satellites in view Keep the device under open sky to acquire GPS Signals VI x Memory Card Almost...

Страница 35: ...Warranties on sensors assume normal usage and will be rendered invalid if the sensors have been exposed to excessive concentrations of gas extended periods of exposure to gas or have been exposed to p...

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ...l 41 41 541 5877 Fax 41 41 541 5878 Email General Information Sales info kesn ch Accounts Department accounts kesn ch USA O ce 233 Rogue River Highway 425 Grants Pass Oregon 97527 5429 USA Tel 1 503 2...

Отзывы: