Gas-Tec
S�ecification
S�ecification
27
IX Specification
Description
Hydrocarbon gas detector
Sensor
Flame Ionisation Detection (FID)
Gas
Calibrated to Methane (CH
4
)
Other correction factors available
Range
0-100 ppm
0-1000 ppm
0-10,000 ppm
Accuracy
+/- 10%
Response time
<2 seconds
H2 Cylinder Life
UK & EU 400 ml = 60 hrs
EU 100 ml = 15 hrs
USA 300 ml = 45 hrs
Case
High impact resistant UV resistant glass reinforced nylon
Size
385 x 205 x 125 mm (15 .2 x 8 x 5 inches)
Weight
2 .4 kg without cylinder
Ingress protection
IP54
Audible alarm
Rising tick increasing with presence of gas
Visual alarm
Indication of status on LCD display
Backlight
Adjustable intensity and time off delay
Battery type
Rechargeable Lithium-ion
Users may not change the battery of this product: it must be replaced at Crowcon
or a service centre .
Battery life
22 hrs
Recharge time
2 .5 hrs
Operating temperature range
-5°C to +55°C
(23°F to 131°F)
Humidity
5 to 95% RH non-condensing
Sample input
Via quick connect filtered coupling
Outputs
Headphone, USB and Trigger sockets
Data logging
5760 datalogs
Event log
5760 events
GPS
Accurate to within 10 m with 5 satellite lock
EMC & approvals
EN50270 & EN50271:2002
Industry approvals
FCC and CE
This device complies with part 15 of the FCC Rules . Operation is
subject to the following two conditions (1) This device may not cause
harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation .
Calibration period
Recommended every 6 months
Warranty
12 months against faulty components or manufacture