2.5

Installation / Operation -

2.5.1

Initial Operation -

1.) If different from factory settings (Table 2.2), set on-card jumpers as desired  (Figure 2.4)
2.) Connect the wall power supply to the 2099 and the wall power supply to 115 VAC, 60 Hz (Fig. 2.1)
3.) Connect  a 1 V P-P 10 MHz signal to J1,

10 MHz SIGNAL IN

 (Figure 2.1, Figure 2.3)

4.) Be sure S1 is in the center (AUTO) position (Figure 2.2)
5.) Be sure DS1 and DS4 are off  and DS2 and DS3 are green (Figure 2.2).
6.) If needed, R20 can be adjusted (Figure 2.2) to SLIGHTLY adjust the internal VCXO frequency with 
S1 in the INTERNAL REF position and monitoring TP1. ALWAYS be sure S1 is in AUTO (center) for 
normal operation.

2.5.2 Switch S1 (Figure 2.2) - S1 selects between Internal Ref / AUTO / External Ref and operates as follows:

External Reference - This is a manual over ride of the AUTO position which forces the 2099-42 to 

always control the internal VCXO by locking it to the external 10 MHz reference. If in this position and the 10 
MHz is removed, the PLL will lose lock and may force the internal VCXO to the high or low rail which causes 
it to be up to plus or minus 1.2 kHz from center frequency. No auto switch over to internal control of the VCXO 
will occur in this mode. The front panel potentiometer, R20 does NOT control the VCXO in this mode. Normal 
operation in this mode will have DS1 red, DS2 and DS3 green, and DS4 off (Figure 2.2).

AUTO - This forces the 2099-42 to always control the internal VCXO by locking it to the external 10 

MHz reference if it is present. If in this position and the 10 MHz is removed, the PLL will lose lock and will 
force the internal VCXO to have the frequency set by front panel potentiometer, R20. R20 DOES control the 
VCXO in this mode if it has automatically switched to Internal. Normal operation in this mode will have DS1 
and DS4 off and DS2 and DS3 green (Figure 2.2).

Internal Reference - This is a manual over ride of the AUTO position which forces the 2099-42 to 

always control the internal VCXO by front panel potentiometer, R20. R20 is a ten turn potentiometer which can 
control the VCXO by up to plus or minus 300 Hz from center frequency. If the 10 MHz external reference is 
present, the PLL Alarm LED, DS4 does not generally light in this mode but will flash occasionally, the rate of 
which is a rough indication of the offset of the VCXO from the reference frequency. The slower the rate of the 
flash, the closer the VCXO frequency is to the 10 MHz reference. No auto switch over to internal control of the 
VCXO will occur in this mode. As mentioned previously, the front panel potentiometer, R20 DOES control the 
VCXO in this mode. Normal operation in this mode will have DS1 red, DS2 red, DS3 green, and DS4 off and 
occasionally red for short flashes (Figure 2.2).

2.5.3 Selecting Clock Frequencies - The output clock frequencies are selected by the jumper on JP4 for 

Clock A and JP5 for Clock B. Setting this jumper closest to the front of the PCB selects 256, then moving it to 
the rear selects 192, 128, 96 and 64. The jumper position closest to the back of the PCB is not used.

2.5.4 Selecting 10 MHz Interface Parameters -  Jumper JP2 selects the input termination of the 

external 10 MHz reference to either 75

 termination (JP2 in DOT, factory setting) or 2 K

 (JP2 in NON-

DOT). 
 

    

                          

2099-42 manual Rev. A

Page 6

10/27/2005

Содержание 2099-42

Страница 1: ...oses of operating the equipment When ordering parts from Cross Technologies Inc be sure to include the equipment model number equipment serial number and a description of the part First Edition Octobe...

Страница 2: ...rials and workmanship for a period of one year after shipment to customer Cross Technologies Inc s obligation under this warranty is limited to repairing or at Cross Technologies Inc s option replacin...

Страница 3: ...manually selects internal or external lock to 10 MHz control of the VCXO Diodes on the DC inputs allow operation from redundant power supplies An open collector indicates Manual operation and relay c...

Страница 4: ...or AUTO select Int Freq Adjusts frequency of internal VCXO when not locked to external 10 MHz INT EXT LED LED lights RED if VCXO controlled internally Green if locked to external 10 MHz MAN LED LED li...

Страница 5: ...nput Output Signals Figure 2 3 shows the input and output signals to the 2099 42 Table 2 1 is a description of the connectors and J4 pin functions 2 4 Jumper Settings Figure 2 4 and Table 2 2 show jum...

Страница 6: ...moved the PLL will lose lock and will force the internal VCXO to have the frequency set by front panel potentiometer R20 R20 DOES control the VCXO in this mode if it has automatically switched to Inte...

Страница 7: ...ll light occasionally depending on how far off in frequency the VCXO is set from locking to it DS3 EXTERNAL REFERENCE PRESENT LED Lights GREEN when 10 MHz external reference is present and RED if belo...

Страница 8: ...RS422 Clock A output 6 NOT USED 7 NOT USED 8 15 VDC Backup In 180 ma max 9 15 VOLTS 180 ma max from J2 Main Power Supply 10 MAJOR Alarm relay output Open or short selectable to pin 12 if PLL alarm or...

Страница 9: ...EQUENCY N A N A N A JUMPER POSITION SELECTS 64 TO 256 CLOCK FREQUENCY CH A JP5 CLOCK A FREQUENCY N A N A N A JUMPER POSITION SELECTS 64 TO 256 CLOCK FREQUENCY CH B TABLE 2 2 JUMPER DESCRIPTIONS K1 R20...

Страница 10: ...and clock B RS422 outputs The external 10 MHz reference frequency also goes to buffer amplifier U7 which drives peak detector CR8 and then goes to op amp U8 the output of which provides a 5 signal if...

Страница 11: ...T MANUAL ALARM UP U3 INT EXT CLOCK PIN 2 PIN 3 VCXO INT FREQ R20 12 288 MHZ VCXO 5V REG VR3 15 A 15B 15 A 15B 12 12 5 12V REG VR1 12V REG VR2 U10 U5 U6 Q2 256 128 64 192 96 INT EXT CLOCK RED GREEN REF...

Отзывы: