![Critikon DINAMAP 8100 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/critikon/dinamap-8100/dinamap-8100_operation-manual_2686191053.webp)
6-4
Figure 6-1. Cover Removal
Changing Fuses
The monitor contains three fuses. Two AC line power fuses
are contained in the power entry module. One battery fuse
is contained on the power supply board and is accessible
only by qualified service personnel.
Replacement of Line Power Fuses
At the rear of the monitor, remove the power cord (if one is
attached). Using a flat bladed tool such as a screwdriver or
key, pry up the lower edge of the door below the power
connector as shown in Figure 6-2 and swing the door
upward on its hinge to reveal the two fuse holders.