3-7
TABLE 3-1 MODEL 8100 FRONT PANEL CONTROLS AND
INDICATORS — Continued
I T EM N A M E
F U N C T I O N
10
PRIOR
DATA
Depression of this switch will cause the
monitor to display the data from the
previous determination and the elapsed
time (in minutes) since the determination
was made. Pressing the switch a second
time will display the data from the next
previous determination, etc.
A total of 100 previous determinations
may be recalled (for up to 99 minutes).
The monitor will exit the PRIOR DATA
display if any other switch is pressed or if
10 seconds elapses with no activity.
Placing the front panel power switch into
the OFF position clears all stored data in
memory.
11
*POWER
ON - OFF
This white pushbutton controls the power
to the monitor. If AC line power is
attached to the rear panel, the battery
charger will operate regardless of
POWER ON/OFF switch position.
12
BATTERY This yellow LED indicates that the
monitor is operating on power from the
internal battery. Flashes when less than
approximately 10% battery charge
remains.
13
*MAINS
AC RE-
CHARGE
This green LED indicates that AC line
power is present and battery is being
charged.
14
CUFF
CONNEC
-TOR
Screw type connectors for connection of
pneumatic (cuff) hose.
* For very old units that have an AC power switch on the rear of the
monitor, the power switch must be in the 'I' position for the statements
in items 11 & 13 to apply.