5-16
Monitor Will Not Power-On
This condition could be caused by a depleted battery.
Check fuse. Attach unit to AC power source. If problem
persists, refer to qualified service personnel.
Monitor Displays Extremely High Readings
This condition could be caused by using a cuff that is too
large or by the cuff being positioned above heart level. Use
the proper size cuff or reposition cuff at heart level. If the
cuff cannot be positioned at heart level, then compensate
by subtracting 1.8 mmHg for every inch below heart level.
Monitor Displays Extremely Low readings
This condition could be caused by using a cuff that is too
large or by the cuff being positioned above heart level. Use
the proper size cuff or reposition cuff at heart level. If the
cuff cannot be positioned at heart level, then compensate
by adding 1.8 mmHg for every inch above heart level.
Patient Alarms Activate After Every Determination
This condition could be caused by narrow high/low alarm
limits. Reset alarm limits as required (see Set Alarm Limits
Procedure, Section 5) and try another determination.