DMS-R-IM
Preventive Maintenance
April 7, 2021
Digital Modular System - Rinse
6-17
6.8
PM For Electrical
NOTE
Electrical enclosure at system rear side of each station (refer
Figure 6-12) is protected by Safety Interlock Panel Switch.
When the enclosure panel is open during system is power
ON, Safety Interlock Door Switch shut off the electrical supply
to the system.
Item
PM Activity/Schedule
Electrical
Check wiring and seek for loosen connections.
HY
(D–daily; W–weekly; M–monthly;, Q–quarterly; HY–half-yearly; Y–yearly; TBD–to be determine)
CAUTION
Power to this system is provided by
DMS-1218: 240VAC / 3-phase / 60Hz / 16Amp
DMS-1826: 240VAC / 3-phase / 60Hz / 31Amp
Disconnect and lockout tag out system prior to any servicing.
Figure 6-13: System Rear View –Safety Panel Switch in electrical enclosure of each station
(pointed by arrow).
Step
Description
1
De-energize system and perform lockout tag out.
2
With help of the electrical schematic, checking wiring and check for
loose wire connection.