oXYgen Scanning Application - User Guide Addendum
9
Functional description of the operating panel
The Power button with LED and the Status indicator with LED are located on
the operating panel.
The LED of the Power
button indicates the Main power switch setting of the
scanner. The LED of the Status indicator indicates the status of the scanner.
NOTE:
Each time you press the Power button, the scanner status toggles between on and off.
The following are the meanings of the LED of the Power button:
•
Off
Power is OFF.
•
Green
Power is ON.
•
Red
Power is OUTSIDE THE TOLERANCE RANGE.
The following are the meanings of the LED of the Status indicator:
•
Orange
Scanner is activated.
•
Red Hardware
problem.
•
Blinking Red
The interlock switches are open.
•
Green
Scanner is idle.
•
Blinking
Green
The scanner is either working or firmware is not
updated. In the event that the firmware is not
updated, you must activate the application version.
•
(Off)
When the Top door is open or the Base glass is not
in position.
NOTE: