16
CPR BAND (Model: CREDO-CB)
CPR Mode
1) Compression Depth Feedback
The LED lights according to the correct compression depth set in the
[Setting] tab. The maximum value is 10 cm and the minimum value is 1
cm of the correct compression depth you can set. The factory-set values
of correct range is 4.5 cm ~ 6.0 cm.
* DC: Depth of Compression
* M: Maximum depth of setting
* m: Minimum depth of setting
* A1:
m
M
m
* A2:
m
M
m
Compression Depth
①
DC < m
Lower 1 orange LED lights up
②
m ≤ DC < A1
1 blue LED lights up
③
A1 ≤ DC < A2 2 blue LEDs light up
④
A2 ≤ DC < M
3 blue LEDs light up
⑤
M ≤ DC
3 blue LEDs and upper Orange LED light
up