![CREATIVE ELECTRONICS SMC-3303X Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/creative-electronics/smc-3303x/smc-3303x_operation-manual_2681426021.webp)
- 18 -
*IDN?
Returns the MODEL identification
RCL<NR1>
Recalls a panel setting
SAVE<NR1>
Saves the panel setting
HELP?
Shows the command list
ERR?
Returns the instrument error messages
5.6 Command Details
Command
ISET<X>:<NR2>
Description
Sets the output current.
Panel operation
Refer to page 11
Response time
Min.70ms
Example
ISET1:2.234
Sets the CH1 output current to
2.234A. (for 1mV 1mA models)
ISET1:2.23
Sets the CH1 output current to
2.23A. (for 10mV 10mA /100mV
10mA models)
Command
ISET<X>?
Description
Returns the output current setting
Response time
Min.70ms
Example
ISET1?
Returns CH1 output current setting.
Command
VSET<X>:<NR2>
Description
Sets the output voltage.
Panel operation
Refer to page 11
Response time
Min.70ms
Example
VSET1:20.345
Sets the CH1 voltage to 20.345V.(for
1mV 1mA models)
Command
VSET<X>?
Description
Returns the output voltage setting.
Response time
Min.70ms
Example
VSET1?
Returns the CH1 voltage setting.
Command
IOUT<X>?
Description
Returns the actual output current.
Response time
Min.70ms
Example
IOUT1?
Returns the CH1 output current.
Command
VOUT<X>?
Description
Returns the actual output voltage.
Response time
Min.70ms
Example
VOUT1?
Returns the CH1 output voltage.
Command
TRACK<NR1>
Description
Selects the operation mode: INDEP,
tracking SER, tracking PAR
Panel operation
Refer to page 12
NR1
0: Independent
1: Tracking series
2: Tracking parallel
Response time
Min.70ms
Command
BEEP<Boolean>
Description
Turns on or off the beeper.
Panel operation
Refer to page 10
Response time
Min.70ms
Example
BEEP1
Turns on the beeper.
Command
OUT<Boolean>
Description
Turns on or off the output.