![CREATIVE ELECTRONICS SMC-3303X Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/creative-electronics/smc-3303x/smc-3303x_operation-manual_2681426020.webp)
- 17 -
5.3 Command Syntax
Command format
1: command header
2: output channel
3: separator
4: parameter
Parameter
Type
<Boolean>
<NR1>
<NR2>
Description
Boolean logic
Integers
Decimal numbers
Example
0 (off), 1 (on)
0, 1, 2, 3
0.1, 3.14, 8.5
Output channel
1 (CH1) or 2 (CH2)
Note
Commands must be capital letters
5.4 Error Messages
The following error messages might appear when the power supply cannot accept the command.
Message contents
Descriptions
a
Program mnemonic too
long
The command length must be 15 characters or less.
b
Invalid character
Invalid characters, such as symbols, are entered. Example: VOUT#
c
Missing parameter
The parameter is missing from the command. Example: VSET: (should
have a number)
d
Data out of range
The entered value exceeds the specification. Example: VSET:33 (should
b
e
≤
32V)
e
Command not allowed
The entered command is not allowed in the circumstance. Example: trying
to set CH2 output while in the tracking mode.
f
Undefined header
The entered command does not exist, or the syntax is wrong.
5.5 Command List
Detailed descriptions of each command starts from the next page.
The “HELP” command shows all the following commands and their meanings, except for the HELP command itself.
Command
Meanings
ISET<X>:<NR2>
Sets the output current
ISET<X>?
Retunes the output current setting
VSET<X>:<NR2>
Sets the output voltage
VSET?
Returns the output voltage setting
IOUT<X>?
Returns the actual output current
VOUT<X>?
Returns the actual output voltage
TRACK<NR1>
Selects the operation mode
BEEP<BOOLEAN>
Turn on or off the beep
OUT<BOOLEAN>
Turn on or off the output
STATUS?
Returns the MODEL status