5. REMOTE CONTROL
5.1 Remote Control Setup
Background
The power supply is capable of being remotely controlled via the USB connection.
Interface
USB slave port, rear panel
COM setting
Set up the COM port inside the PC according to the following list:
Baud rate: 9600
Parity bit: None
Data bit: 8
Stop bit: 1
Data flow control: None
Functionality check
Run this query command via the terminal application such as MTTTY (Multi-threaded
TTY).
*IDN?
This should return the identification information: series number, firmware version.
5.2 Remote Connection Step
Enter the remote control
mode
1.
Connect the USB cable to the slave port.
2.
The connection will be automatically established, and the front panel shows
“USB…YES” message.
3.
The power supply also automatically enters the lock state (the Lock key will be
activated).
Exit the remote control
mode
1.
Disconnect the USB cable from the rear.
2.
The display shows “USB…NO” message.
3.
Unlock the power supply by pressing the Lock key until it turns off backlight.
4.
The power supply goes back to local operation mode.