SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with
caution. The rules listed below must be carefully followed for safe operation.
1.
NEVER
apply voltage or current to the meter that exceeds the
specified maximum:
Input Limits
Function Maximum
Input
V DC 200mV range
250V DC or AC
V DC/V AC >200mV ranges
1000V DC or 700V AC
mA DC/AC
200mA DC/AC
20A DC/AC
20A DC/AC (30 seconds max every 15
minutes)
Frequency, Resistance,
Capacitance, Diode test,
Continuity, Temperature
250V DC/AC
2.
USE EXTREME CAUTION
when working with high voltages.
3.
DO NOT
measure voltage if the voltage on the "COM" input jack
exceeds 500V above earth ground.
4.
NEVER
connect the meter leads across a voltage source while the
function switch is in the current, resistance, or diode mode. Doing so
can damage the meter.
5.
ALWAYS
discharge filter capacitors in power supplies and disconnect
the power when making resistance or diode tests.
6.
ALWAYS
turn off the power and disconnect the test leads before
opening the doors to replace the fuse or batteries.
7.
NEVER
operate the meter unless the back cover and the battery and
fuse doors are in place and fastened securely.