InstroTek
21
An alternative method of updating the calibrations manually
From the Gauge Ready Screen, press
MENU
– Use the Arrow
keys to select ’
13. Calibrations
’.
Press
Enter
. Use the Arrow keys to select ‘
4. Send to USB
’.
Insert a Thumb Drive, and press
ENTER
.
After the data has been downloaded to the Thumb Drive,
remove the Drive and insert it into your PC.
The calibrations will have been saved in the folder
\InstroTek\503 HydroProbe\ CAL 01-01-80_12_00_AM.XML
Open the file with EXCEL, make any changes you want and
save it back to the thumb drive as \InstroTek\503
HydroProbe\LoadCal.XML
Insert the Thumb Drive back into the gauge. Use the Arrow
keys to select ‘
6 Load From USB
’.
Press
Enter
.
Self-Calibration
From the Gauge Ready Screen, press
MENU
– Use the Arrow
keys to select ’
13. Calibrations’
.
Press
Enter
. Use the Arrow keys to select ‘
2. Enter/Edit Cal
’.
Press
Enter
. Enter the
Password
(
3548
).
Use the Arrow keys to select ‘
3. Self-Calibration
’. Use the
Arrow keys to select ‘
2. Gauge Derived
’.
Press
Enter
.