Alarms and Troubleshooting
3-8
Puritan Bennett 540™ Ventilator User’s Manual
CHECK PROXIMAL
LINE1
*IF PERSISTS
REPLACE VENT
1. Loss of signal from the
proximal pressure sensor
Consequence: switch to internal
pressure sensor for the pressure
measurement.
Alarm activation occurs:
•
In the event of signal loss (1):
After one ventilation cycle
or
In the event of signal loss(2) and
after the 17
th
breath cycle:
After 17 seconds for PRES A/C
and VOL A/C modes, or after the
maximum time between 17
seconds and Apnea Time + 4
seconds for CPAP/PSV/ P SIMV,
and V SIMV modes
MP
Yes
No
CHECK PROXIMAL
LINE2
*IF PERSISTS
REPLACE VENT
Proximal pressure < 0.6 cmH2O
for 100 ms during inspiration
phase of 2nd breath cycle
Ventilator response: Switch to
internal pressure sensor for
pressure measurement.
MP
Yes
No
CHECK/CHANGE EXH
VALVE2
Abnormally high expired flow
during the inspiratory phase of
three consecutive breaths (in
double-limb setup).
Alarm activation occurs:
After three consecutive breaths.
MP
Yes
No
CHECK
SETTINGS
Alarm activation occurs:
•
Systematically after software
versions have changed.
•
Loss of memorized
parameters
Consequence:
•
Locking Key disabled
•
Out-of-range settings are
replaced by their default
values
MP
Yes
Yes
CHK EXH VALV
RESTART VENT*
*IF PERSISTS
REPLACE VENT
•
Faulty pressure valve sensor
signal at ventilation start.
HP Yes
Yes
Table 3-1.
Overview of Alarms (Continued)
Alarm Message
Cause/Ventilator Response
Priority
Audio
Paused
Avail.
Alarm
Paused
Avail.
Содержание Puritan Bennett 540
Страница 1: ...Puritan Bennett 540 Ventilator User s Manual...
Страница 6: ...This page intentionally blank...
Страница 8: ...This page intentionally blank...
Страница 10: ...This page intentionally blank...
Страница 80: ...Installation and Assembly 4 20 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 86: ...Operating Procedures 5 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 94: ...This page intentionally blank...
Страница 100: ...Routine Maintenance 8 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Страница 124: ...Unpacking and Preparation C 4 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Страница 140: ......