5
1.3.5
Transferring the Patient
WARNING
When the patient is elevated a few inches off a stationary surface and before moving
the patient, check again to make sure that all hardware or straps are properly
connected. If any attachment is NOT properly in place, lower the patient back onto
the bed and rectify the issue.
Wheelchair wheels MUST be locked before lowering the patient into the wheelchair
for transport.
Make sure the wheelchair weight capacity can withstand the patient's weight prior
to loading the patient.
Be sure to check the sling attachments each time the sling is removed and replaced,
to ensure that it is properly attached before the patient is removed from the bed or
chair.
Mast pivot under the rubber boot must be tight to ensure safe use of the patient lift.
Bolt must be checked at least every six months in conjunction with periodic
maintenance.
1.3.6
Performing Maintenance
WARNING
ONLY qualified personnel should carry out the lift maintenance.
DO NOT attempt to open the pre-sealed hydraulic pump or obtain local service, for
it will VOID the warranty and might result in damage. Consult your dealer or
manufacturer for further information.
After the first year of operation, the hooks of the swivel bar and the mounting
brackets of the boom should be inspected every three months to determine the
extent of wear. If these parts become worn, replacement must be made.
After the first year of operation, inspect the swivel bar and the eye of the boom (to
which it attaches) for wear. If the metal is worn, the parts MUST be replaced. Make
this inspection every six months thereafter.
Inspect the casters and axle bolts every six months to check for tightness and wear.