15
4.1.2
Raising/lowering the lift
WARNING
The rear casters of the patient lift should not be locked when lifting an individual.
Locked casters could cause the lift to tip and endanger the patient and assistants. It
is recommended that the rear casters be left unlocked during lifting procedures to
allow the patient lift to stabilize itself when the patient is initially lifted from a chair,
bed or any stationary object.
I.
To raise the patient, first make sure the control valve is at close position (Valve
handle parallel to the pump handle).
II.
Maneuver the pump handle in an up and down motion to lift up the patient.
III.
To lower the patient, turn the control valve away from the pump hand in the open
position. The weight of the patient will lower the lift. The degree of control valve
opening will vary the speed of descent.
The safety mechanism built into the lift allows the patient to descend slowly even
when the control valve is fully open.