- 5 -

 

 

No. Name 

Functions 

AT-2G Gas Probe 

Gas inlet. 

DF-114 Drain Filter 

Protects the Gas Leak Detector from water, dust, 

and other foreign matter. 

FE-106 Dust Filter 

Protects the Gas Leak Detector from fine dust, 

sand, and other foreign matter. 

Outlet 

Gas outlet to exhaust sampled gas. 

POWER/ SENSITIVITY Button 

Turns the power ON and OFF. 

Sets the detection sensitivity level. 

BUZZER Button 

Turns the alarm buzzer and operation sound ON 

and OFF.   

Switches the gas types for two-gas model. 

ZERO/ RESTART PUMP/ 

BACKLIGHT Button 

Restarts pump operation after a pump error.

 

Also 

lights the LCD backlight and displays the target gas 

name on the LCD. 

Alarm/ Error Lamps (red) 

Light

 

up or flash when detecting gas leakage.

 

The 

flashing frequency increases as the concentration 

of the gas leak increases. The lamps light if the 

concentration is very high.

 

These lamps also flash 

when an error occurs in the pump or sensor. 

Buzzer hole 

The buzzer sounds. 

10 

LCD  

 

 

 

 

 

 

 

Displays the gas sensitivity and messages. 

(a)

 

The remaining battery charge is displayed. 

Battery charge: 

High . . . . . . . . . . . . . . . . . Low 

 

 

 

(b)

 

It indicates a possible status of zero adjustment.

 

(c)

 

It shows the operation sound. (The indicator is 

displayed only when the operation sound is 

turned ON.) 

(d) It shows the alarm sound.

 

(The icon is displayed 

only when the alarm sound is ON.) 

(e) The pump operation is displayed.

 

(The indicator 

rotates if the pump is operating normally.)   

(f)

 

The detection sensitivity level, target gas name, 

and errors are displayed. 

11 

Gas Sensitivity Lamps (green) 

Indicate the detection sensitivity level.

 

There are 

five detection sensitivity levels. 

12 

Battery Lid 

The cover to the battery compartment. 

 

(a) 

(b) 

(c) 

(d) 

(f) 

(e) 

 

- 4 -

 

 

2. Part Names and Functions 

 

 

 

 

 

 

11 

10 

12 

Содержание XP-704 III

Страница 1: ...Keep this instruction manual available for quick reference when needed Read this instruction manual and thoroughly understand before using the product HCFC Detector...

Страница 2: ...11 Deactivating the Alarm Buzzer 12 Lighting the Backlight 12 4 Error Displays 13 5 Replacement of Consumable Parts 14 Filter Element Replacement 14 Dust Filter Replacement 15 Battery Replacement 16 6...

Страница 3: ...ew Cosmos distributor Replacements for any missing or damaged items will be sent Options Sold Separately 1 1 Refer to page 18 for details of options Name Quantity Gas Leak Detector with soft case 1 DF...

Страница 4: ...horoughly this Instruction Manual regardless of your experience with using Gas Leak Detectors Do not use this Gas Leak Detector for improper purpose Do not use this Gas Leak Detector in undocumented w...

Страница 5: ...eplace batteries if the batteries are leaking when the remaining battery charge is low To avoid damaging its performance never attempt to disassemble alter or change the structure or electric circuits...

Страница 6: ...4 2 Part Names and Functions 8 1 2 11 7 9 5 4 10 6 3 12...

Страница 7: ...hing frequency increases as the concentration of the gas leak increases The lamps light if the concentration is very high These lamps also flash when an error occurs in the pump or sensor 9 Buzzer hol...

Страница 8: ...6 No Name Functions 1 Side Zipper Open the zipper to replace the batteries 2 Back Pocket The Operating Guide is stored here 3 Inside Pocket A Replacement Filter Element is stored here Soft case 1 3 2...

Страница 9: ...up both sides and close the Velcro The Gas Leak Detector is in the soft case at the time of shipment from the factory 2 Insert AT 2 Gas Probe to DF 114 Drain Filter AT 2 Gas Probe is inserted to DF 11...

Страница 10: ...nsitivity level 3 Before warm up is completed the highest sensitivity level 0 lamp will flash The maximum warm up is three minutes Always perform a daily check before using the Gas Leak Detector Refer...

Страница 11: ...ion sensitivity level is decreased beyond 1 the level will return to level 5 To Increase the Detection sensitivity level Each time pressing the POWER SENSITIVITY button twice like double click the buz...

Страница 12: ...ep sound and then returns to the detection sensitivity level indication When zero adjustment is not available the buzzer will sound and the detection sensitivity level remains displaying on the LCD 4...

Страница 13: ...ode When it enters automatic gas exhaust mode the buzzer will sound once 60 will be displayed on the LCD and a countdown will start 59 58 Silent Mode No Operation Sound The alarm buzzer can be turned...

Страница 14: ...d will be turned ON or OFF The alarm sound icon on the LCD will light buzzer ON or go out buzzer OFF Lighting the Backlight Tap the ZERO RESTART PUMP BACKLIGHT button to light up the backlight for 30...

Страница 15: ...g ON the power several times there may be a problem with the sensor Request repairs Pump Error Error lamps flash Water or liquid may have been sampled or the inlet may be blocked Remove the water or l...

Страница 16: ...unction properly and will not be able to detect correctly Attach the filter case securely If the filter case is loose water or other foreign matter may leak into the Gas Leak Detector If water or liqu...

Страница 17: ...Turn the cap of the Drain Filter counterclockwise to remove together with the Gas Probe 2 Remove FE 106 Dust Filter and replace with a new filter 3 Attach the Drain Filter and reassemble Make sure to...

Страница 18: ...en the side zipper 2 Open the battery lid of the back of the Gas Leak Detector and install two new batteries size AA alkaline as shown on the bottom of the compartment 3 Close the battery lid firmly 4...

Страница 19: ...the alarm sound icon is being displayed and the alarm lamp red does not flash or the alarm buzzer does not sound contact your New Cosmos distributor Gas Probe Check the Gas Probe for any wear or damag...

Страница 20: ...ng and coupling Gas Collector AT 5 20524430 For tubing and coupling Annealed Copper Tube AT 11 20524431 59060000 with tube For tubing secluded locations or devices Sampling Hose SH 4 05 20524531 Used...

Страница 21: ...will not turn ON when the POWER SENSITIVY button is pressed The battery polarity is reversed Insert the batteries correctly Battery Replacement on page 16 Battery exhaustion Replace the batteries Disp...

Страница 22: ...After examination of allegedly defective product return to New Cosmos with freight prepaid should the product fail to conform to this warranty customer s only remedy and New Cosmos s only obligation...

Страница 23: ...g Pressure Range Atmospheric pressure 800 to 1100 hPa Operating Temperature and Humidity Range 0 to 40 degrees C Less than 85 RH with no condensation Dimensions Approx 38 W x 130 H x 32 D mm excluding...

Страница 24: ...cted of a metal oxide semiconductor that is sintered in a ball on a platinum wire coil that functions as both a heater and electrode When combustible gas is absorbed by the surface of the metal oxide...

Страница 25: ...23 MEMO...

Страница 26: ...MEMO...

Страница 27: ...MEMO...

Страница 28: ...naka Yodogawa ku Osaka 532 0036 Japan Tokyo Branch New Cosmos Electric Co Ltd 2 6 2 Hamamatsu cho Minato ku Tokyo 105 0013 Japan Phone 81 3 5403 2715 Fax 81 3 5403 2710 Email e info new cosmos co jp U...

Отзывы: