Dust Filter Replacement 

 

 

 

 

 

 

 

 

 

FE-106 Dust Filter prevents to enter dust particles and sand into the pump inside the 

Gas Leak Detector. Do not remove the Dust Filter.   

The Dust Filter will be kept clean in normal use (with DF-114 Drain Filter). However, 

the Dust Filter is available for replacement as needed (sold separately). 

 

 

1)  Turn the cap of the Drain Filter counterclockwise 

to remove together with the Gas Probe. 

 

 

 

 

 

 

 

 

 

2)  Remove FE-106 Dust Filter and replace with a 

new filter. 

 

 

 

 

 

 

 

 

 

 

 

 

3)  Attach the Drain Filter and reassemble. 

  Make sure to attach FE-2 Filter Element in the Drain Filter when using 

FE-106 Dust Filter. 

  It cannot prevent water from entering into the Gas Leak Detector 

unless using FE-2 Filter Element. 

  Do not replace FE-106 Dust Filter while the Gas Leak Detector is 

operating. 

CAUTION

 

Gas Probe   

Cap   

Drain Filter

 

Remove 

Insert 

FE-106 Dust Filter  

 

- 14 -

 

5. Replacement of Consumable Parts 

 

 

  

Filter Element Replacement 

 

When the Filter Element is dirty or wet, or when the water or liquid is inside the Drain 

Filter, clean out the Drain Filter and replace the Filter Element. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1)  Turn the filter case counterclockwise and 

remove it. 

 

 

 

 

 

 

2)  Remove the O-ring from inside the cap. 

 

3)  Replace the Filter Element with a new filter. 

 

 

 

 

 

 

 

 

 

 

 

 

4)  Attach the O-ring, put the filter case back 

together and reassemble. 

 

 Do not replace the Filter Element while the Gas Leak Detector is 

operating.

 

Dust or other foreign matter may enter the Gas Leak 

Detector. 

  Mount the Filter Element correctly. If the Filter is not in place properly, it 

will not function properly and will not be able to detect correctly. 

  Attach the filter case securely.

 

If the filter case is loose, water or other 

foreign matter may leak into the Gas Leak Detector. 

  If water or liquid reaches inside the Gas Leak Detector through the Filter 

Element, it is in need of repair.

 

It will not be able to detect correctly. 

CAUTION

 

Do not push or poke the Filter Element. It may damage the waterproof structure. 

NOTE 

O-ring

FE-2  

Filter Element 

Filter Case

Cap

Drain Filter 

If water, dust, or other foreign 

matter is in the Drain Filter, be 

careful not to enter into the Gas 

Leak Detector. 

CAUTION

 

Clean the inside of the cap and filter 

case with a dry cloth as needed.

 

R

emove water or liquid completely and 

dry the parts before replacing with a 

new Filter Element. 

NOTE 

Make sure to attach FE-2 Filter 

Element between the cap and the 

filter case. 

CAUTION

 

Содержание XP-704 III

Страница 1: ...Keep this instruction manual available for quick reference when needed Read this instruction manual and thoroughly understand before using the product HCFC Detector...

Страница 2: ...11 Deactivating the Alarm Buzzer 12 Lighting the Backlight 12 4 Error Displays 13 5 Replacement of Consumable Parts 14 Filter Element Replacement 14 Dust Filter Replacement 15 Battery Replacement 16 6...

Страница 3: ...ew Cosmos distributor Replacements for any missing or damaged items will be sent Options Sold Separately 1 1 Refer to page 18 for details of options Name Quantity Gas Leak Detector with soft case 1 DF...

Страница 4: ...horoughly this Instruction Manual regardless of your experience with using Gas Leak Detectors Do not use this Gas Leak Detector for improper purpose Do not use this Gas Leak Detector in undocumented w...

Страница 5: ...eplace batteries if the batteries are leaking when the remaining battery charge is low To avoid damaging its performance never attempt to disassemble alter or change the structure or electric circuits...

Страница 6: ...4 2 Part Names and Functions 8 1 2 11 7 9 5 4 10 6 3 12...

Страница 7: ...hing frequency increases as the concentration of the gas leak increases The lamps light if the concentration is very high These lamps also flash when an error occurs in the pump or sensor 9 Buzzer hol...

Страница 8: ...6 No Name Functions 1 Side Zipper Open the zipper to replace the batteries 2 Back Pocket The Operating Guide is stored here 3 Inside Pocket A Replacement Filter Element is stored here Soft case 1 3 2...

Страница 9: ...up both sides and close the Velcro The Gas Leak Detector is in the soft case at the time of shipment from the factory 2 Insert AT 2 Gas Probe to DF 114 Drain Filter AT 2 Gas Probe is inserted to DF 11...

Страница 10: ...nsitivity level 3 Before warm up is completed the highest sensitivity level 0 lamp will flash The maximum warm up is three minutes Always perform a daily check before using the Gas Leak Detector Refer...

Страница 11: ...ion sensitivity level is decreased beyond 1 the level will return to level 5 To Increase the Detection sensitivity level Each time pressing the POWER SENSITIVITY button twice like double click the buz...

Страница 12: ...ep sound and then returns to the detection sensitivity level indication When zero adjustment is not available the buzzer will sound and the detection sensitivity level remains displaying on the LCD 4...

Страница 13: ...ode When it enters automatic gas exhaust mode the buzzer will sound once 60 will be displayed on the LCD and a countdown will start 59 58 Silent Mode No Operation Sound The alarm buzzer can be turned...

Страница 14: ...d will be turned ON or OFF The alarm sound icon on the LCD will light buzzer ON or go out buzzer OFF Lighting the Backlight Tap the ZERO RESTART PUMP BACKLIGHT button to light up the backlight for 30...

Страница 15: ...g ON the power several times there may be a problem with the sensor Request repairs Pump Error Error lamps flash Water or liquid may have been sampled or the inlet may be blocked Remove the water or l...

Страница 16: ...unction properly and will not be able to detect correctly Attach the filter case securely If the filter case is loose water or other foreign matter may leak into the Gas Leak Detector If water or liqu...

Страница 17: ...Turn the cap of the Drain Filter counterclockwise to remove together with the Gas Probe 2 Remove FE 106 Dust Filter and replace with a new filter 3 Attach the Drain Filter and reassemble Make sure to...

Страница 18: ...en the side zipper 2 Open the battery lid of the back of the Gas Leak Detector and install two new batteries size AA alkaline as shown on the bottom of the compartment 3 Close the battery lid firmly 4...

Страница 19: ...the alarm sound icon is being displayed and the alarm lamp red does not flash or the alarm buzzer does not sound contact your New Cosmos distributor Gas Probe Check the Gas Probe for any wear or damag...

Страница 20: ...ng and coupling Gas Collector AT 5 20524430 For tubing and coupling Annealed Copper Tube AT 11 20524431 59060000 with tube For tubing secluded locations or devices Sampling Hose SH 4 05 20524531 Used...

Страница 21: ...will not turn ON when the POWER SENSITIVY button is pressed The battery polarity is reversed Insert the batteries correctly Battery Replacement on page 16 Battery exhaustion Replace the batteries Disp...

Страница 22: ...After examination of allegedly defective product return to New Cosmos with freight prepaid should the product fail to conform to this warranty customer s only remedy and New Cosmos s only obligation...

Страница 23: ...g Pressure Range Atmospheric pressure 800 to 1100 hPa Operating Temperature and Humidity Range 0 to 40 degrees C Less than 85 RH with no condensation Dimensions Approx 38 W x 130 H x 32 D mm excluding...

Страница 24: ...cted of a metal oxide semiconductor that is sintered in a ball on a platinum wire coil that functions as both a heater and electrode When combustible gas is absorbed by the surface of the metal oxide...

Страница 25: ...23 MEMO...

Страница 26: ...MEMO...

Страница 27: ...MEMO...

Страница 28: ...naka Yodogawa ku Osaka 532 0036 Japan Tokyo Branch New Cosmos Electric Co Ltd 2 6 2 Hamamatsu cho Minato ku Tokyo 105 0013 Japan Phone 81 3 5403 2715 Fax 81 3 5403 2710 Email e info new cosmos co jp U...

Отзывы: