- 17 -

 

 

6. Maintenance Check 

 

 

The Gas Leak Detector is precision measuring equipment. Maintenance in accordance 

with the following items is imperative for maintaining performance of the Gas Leak 

Detector and ensuring safety.   

In case of dropping the Gas Leak Detector or being wet by using under any conditions 

other than the normal environment (refer to 

Precautions 

in

 1. Introduction

 on page 3 and 

Range

 in

 9. Specifications 

on page 23), perform a daily inspection or request a periodic 

inspection (will be charged). 

 

  

Daily Check 

 

Items to Check 

Description 

Alarm Function 

Remove the Dust Filter (refer to 

Dust Filter Replacement

 on page 

15), attach the Drain Filter and turn ON the power. Then set the 

detection sensitivity level to 4 and confirm that the alarm lamp (red) 

flashes and the alarm buzzer will sound when suctioning inspection 

gas for 0.5 sec. 

After inspection, turn OFF the power and make sure to attach the 

Dust Filter. 

When the alarm sound icon is being displayed and the alarm lamp 

(red) does not flash or the alarm buzzer does not sound, contact 

your New Cosmos distributor. 

Gas Probe 

Check the Gas Probe for any wear or damage which would affect 

operations.

 

If any, replace the Gas Probe. 

Filter Element 

When the Drain Filter is dirty or wet, or the water or liquid is in the 

Drain Filter, clean out the Drain Filter and replace the Filter Element.

 

(Refer to 

Filter Element Replacement

 on page 14.) 

Battery charge 

If the remaining battery charge is low, replace the batteries.

 

(Refer 

to

 Battery Replacement

 on page 16.) 

Airtightness 

Turn ON the power and confirm that "E-P" (refer to 

Pump Error

 on 

page 13) is displayed on the LCD by blocking the inlet of the Gas 

Probe. 

If an error occurs, the airtightness is normal. Press the restart pump 

button. 

If an error does not occur, the Drain Filter is damaged or the airtight 

of diaphragm is damaged. Replace the parts or request repairs. 

 

 

 

  

Periodic Inspection 

 

To maintain the accuracy of the Gas Leak Detector, request a periodic inspection at your 

New Cosmos distributor or directly from New Cosmos at least once a year. 

 

- 16 -

 

  

Battery Replacement 

 

 

 

 

 

 

When the remaining battery charge indicator changes to            , the battery charge is 

low.  

When the battery voltage drops and reaches to the final voltage, “E-B” (refer to 

Battery 

Alarm

 on page 13) will be displayed on the LCD and the Gas Leak Detector will not be 

working. Replace the batteries. 

 

 

 

 

 

 

 
1)  Open the Velcro tape of the soft case and open the side zipper. 
2)  Open the battery lid of the back of the Gas Leak Detector and install two new 

batteries (size AA alkaline) as shown on the bottom of the compartment. 

3)  Close the battery lid firmly. 
4)  Zip the right and left of the soft case and close the Velcro tape. 

 

 

 

 

 

  Be sure to use two AA alkaline dry cell batteries (LR6).   

WARNING

 

  Use the same kind of unused batteries. 

  Always replace two batteries at the same time. 

 

Due to the characteristics of the batteries, the battery life will be shorter in 

low-temperature environment.

 

NOTE 

Battery lid 

Velcro tape 

Содержание XP-704 III

Страница 1: ...Keep this instruction manual available for quick reference when needed Read this instruction manual and thoroughly understand before using the product HCFC Detector...

Страница 2: ...11 Deactivating the Alarm Buzzer 12 Lighting the Backlight 12 4 Error Displays 13 5 Replacement of Consumable Parts 14 Filter Element Replacement 14 Dust Filter Replacement 15 Battery Replacement 16 6...

Страница 3: ...ew Cosmos distributor Replacements for any missing or damaged items will be sent Options Sold Separately 1 1 Refer to page 18 for details of options Name Quantity Gas Leak Detector with soft case 1 DF...

Страница 4: ...horoughly this Instruction Manual regardless of your experience with using Gas Leak Detectors Do not use this Gas Leak Detector for improper purpose Do not use this Gas Leak Detector in undocumented w...

Страница 5: ...eplace batteries if the batteries are leaking when the remaining battery charge is low To avoid damaging its performance never attempt to disassemble alter or change the structure or electric circuits...

Страница 6: ...4 2 Part Names and Functions 8 1 2 11 7 9 5 4 10 6 3 12...

Страница 7: ...hing frequency increases as the concentration of the gas leak increases The lamps light if the concentration is very high These lamps also flash when an error occurs in the pump or sensor 9 Buzzer hol...

Страница 8: ...6 No Name Functions 1 Side Zipper Open the zipper to replace the batteries 2 Back Pocket The Operating Guide is stored here 3 Inside Pocket A Replacement Filter Element is stored here Soft case 1 3 2...

Страница 9: ...up both sides and close the Velcro The Gas Leak Detector is in the soft case at the time of shipment from the factory 2 Insert AT 2 Gas Probe to DF 114 Drain Filter AT 2 Gas Probe is inserted to DF 11...

Страница 10: ...nsitivity level 3 Before warm up is completed the highest sensitivity level 0 lamp will flash The maximum warm up is three minutes Always perform a daily check before using the Gas Leak Detector Refer...

Страница 11: ...ion sensitivity level is decreased beyond 1 the level will return to level 5 To Increase the Detection sensitivity level Each time pressing the POWER SENSITIVITY button twice like double click the buz...

Страница 12: ...ep sound and then returns to the detection sensitivity level indication When zero adjustment is not available the buzzer will sound and the detection sensitivity level remains displaying on the LCD 4...

Страница 13: ...ode When it enters automatic gas exhaust mode the buzzer will sound once 60 will be displayed on the LCD and a countdown will start 59 58 Silent Mode No Operation Sound The alarm buzzer can be turned...

Страница 14: ...d will be turned ON or OFF The alarm sound icon on the LCD will light buzzer ON or go out buzzer OFF Lighting the Backlight Tap the ZERO RESTART PUMP BACKLIGHT button to light up the backlight for 30...

Страница 15: ...g ON the power several times there may be a problem with the sensor Request repairs Pump Error Error lamps flash Water or liquid may have been sampled or the inlet may be blocked Remove the water or l...

Страница 16: ...unction properly and will not be able to detect correctly Attach the filter case securely If the filter case is loose water or other foreign matter may leak into the Gas Leak Detector If water or liqu...

Страница 17: ...Turn the cap of the Drain Filter counterclockwise to remove together with the Gas Probe 2 Remove FE 106 Dust Filter and replace with a new filter 3 Attach the Drain Filter and reassemble Make sure to...

Страница 18: ...en the side zipper 2 Open the battery lid of the back of the Gas Leak Detector and install two new batteries size AA alkaline as shown on the bottom of the compartment 3 Close the battery lid firmly 4...

Страница 19: ...the alarm sound icon is being displayed and the alarm lamp red does not flash or the alarm buzzer does not sound contact your New Cosmos distributor Gas Probe Check the Gas Probe for any wear or damag...

Страница 20: ...ng and coupling Gas Collector AT 5 20524430 For tubing and coupling Annealed Copper Tube AT 11 20524431 59060000 with tube For tubing secluded locations or devices Sampling Hose SH 4 05 20524531 Used...

Страница 21: ...will not turn ON when the POWER SENSITIVY button is pressed The battery polarity is reversed Insert the batteries correctly Battery Replacement on page 16 Battery exhaustion Replace the batteries Disp...

Страница 22: ...After examination of allegedly defective product return to New Cosmos with freight prepaid should the product fail to conform to this warranty customer s only remedy and New Cosmos s only obligation...

Страница 23: ...g Pressure Range Atmospheric pressure 800 to 1100 hPa Operating Temperature and Humidity Range 0 to 40 degrees C Less than 85 RH with no condensation Dimensions Approx 38 W x 130 H x 32 D mm excluding...

Страница 24: ...cted of a metal oxide semiconductor that is sintered in a ball on a platinum wire coil that functions as both a heater and electrode When combustible gas is absorbed by the surface of the metal oxide...

Страница 25: ...23 MEMO...

Страница 26: ...MEMO...

Страница 27: ...MEMO...

Страница 28: ...naka Yodogawa ku Osaka 532 0036 Japan Tokyo Branch New Cosmos Electric Co Ltd 2 6 2 Hamamatsu cho Minato ku Tokyo 105 0013 Japan Phone 81 3 5403 2715 Fax 81 3 5403 2710 Email e info new cosmos co jp U...

Отзывы: