10 ALARMS IN PORTAL
The CORSANO TRIALS app collects metrics from the device and synchronizes to the
cloud. When alarm conditions are met, data are highlighted in the patient summary page,
available for the physicians to check the patients.
The delay in alarm generation depends on the synchronization time of the device data
with the cloud. When the connection from the device up to the cloud is maintained, this
can take up to 10 minutes. The data and alarm indication are available for the physician
on the Corsano portal.
On the portal, the physicians can define the alarm levels conditions. The delay to trigger
the alarm condition depends on the synchronization time of the data collection chain and
can go up to 10 minutes. More if the data transmission link is broken.
Alarm limits can be defined for the following metrics:
•
Heart rate
•
Temperature
•
Respiratory rate
•
SpO2
Caution: The alarms are information to help the physicians to see if the metrics are in a
pre-defined range. The system is not intended to be used in high acuity environments
where the condition of the patient may require immediate action.