The CardioWatch 287-2B is not intended for infants weighing less than 10 kg.
The CardioWatch 287-2B is not able to display pacemaker pulses.
The Heart Rate from ECG is calculated by counting the number of R peaks in 6 seconds
and multiply by 10.
A pause in ECG is determined when baseline is stabilized and there is no R peak for more
than 3 seconds.
The CardioWatch 287-2B can measure ST segment shifts:
•
Analysis is performed on the single lead
•
No operator selectable detection criteria
•
Information is displayed episode by episode
•
Heart rate and displacement are reported for each episode