background image

N62 Piezo Nanopositioning Scanner User Manual

6

4. Operating environment

   A clean, dustfree and non-corrosive environment

   Operating temperature : -20~80℃

   Relative humidity : <60%

   Recommended voltage range is 0~120V for long-term use and lifetime

FIG.2 N62 Installation diagram

5. Installation and operation

Safety check should be carried out before installation, such as whether the connection wire 

is damaged or broken, whether the drive power supply is grounded, etc.

5.1 Installation and Precautions 

Installation diagram of N62 is as follows:

1.Ensure that the installation plane is relatively flat.

2.Use M3 screws to mount N62 on the mounting surface directly and tighten it.

3.Grounding protection for connection, as shown below.

M3 hole for load mounting

M2.5 hole for grounding

Motion surface 

M3 hole for mounting

Содержание N62

Страница 1: ...N62 Piezo Nanopositioning Scanner User Manual Version V1 0 This document describes the following products N62 XY20S N62 XY20S...

Страница 2: ...If you disassemble or transform the product without permission our company will not be responsible for any consequences arising therefrom To change the product model and other false sales of our compa...

Страница 3: ...5 4 Operating environment 6 5 Installation and operation 6 5 1 Installation and Precautions 6 5 2 Load 7 5 3 Operating 10 6 Transportation and inspection 11 6 1 Transportation 11 6 2 Open Box Inspecti...

Страница 4: ...tion and use 1 2 Symbol and meaning DANGER Improper handling may could cause injury to the operator Attention Improper handling can cause damage to the equipment No order between items 1 3 Users of Th...

Страница 5: ...casually if you need a longer cable please contact CoreMorrow A change in temperature and pressure will charge the piezoelectric actuator and the actuator will remain energized for a period of time wh...

Страница 6: ...hanism has advantages of no friction high positioning accuracy Flexible hinge is an element not affected by static and dynamic friction based on the elastic deformation of metal material e g steel the...

Страница 7: ...n biotechnology interferometry quality detection optical detection nanometer positioning microscopy and imaging image processing and other fields with a good development prospect Y FIG 1 The Motion Di...

Страница 8: ...y check should be carried out before installation such as whether the connection wire is damaged or broken whether the drive power supply is grounded etc 5 1 Installation and Precautions Installation...

Страница 9: ...e connector with the corresponding controller according to the identification The wiring mode of LEMO connector is shown as below Safety washer Flat washer Platform Cable connector Ground protection S...

Страница 10: ...of the product we do not recommend that the load is too high If a high load must be used please appropriately low the position of the center of gravity of the load The wrong mounting way is shown in t...

Страница 11: ...expose the product and scratch the product affecting the motion accuracy of the product Short bolts will lead to insufficient compactness of the connection and easy looseness which will damage the int...

Страница 12: ...controller operations can be carried out 1 Please read the user manual of the controller carefully before operation and then start the controller 2 Set the controller output voltage to zero first 3 Ch...

Страница 13: ...ge shall not exceed the voltage 150V range of N62 If any abnormal noise or oscillation occurs during use turn off the power immediately and check the parameter setting 6 2 Open Box Inspection Inspecti...

Страница 14: ...nevenness Mount the platform on a flat surface Loose connection Fasten connection Wrong connection Check if the wiring is correct Vibration or inaccurate positioning at startup Incorrect input control...

Страница 15: ...oduct from the controller before cleaning the surface of the product When cleaning gently wipe the surface of the product with a towel dipped in a small amount of detergent and then clean it with alco...

Страница 16: ...te The tolerances of the above data are all 20 8 2 Drawing FIG 10 Drawings Models N62 XY20S K Unit Active axis XY Travel range 0 120V 8 m Travel range 0 150V 10 m Unloaded resonant frequency 2000 Hz L...

Страница 17: ...250 CTL as shown below The sensor connector adopts the four core lemo model LEMO ERA 0s 304 CLL as showne below Note Open loop products have only PZT drive line no sensor line FIG 13 Definition of Se...